Teledyne 3000PA operating instructions Index, Removing the Micro-Fuel Cell

Page 53

Percent Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

3.Divide the raw output reading by the percent oxygen concentration of your span gas.

If the quotient is less than the Index value for the cell class you are using, replace the cell.

Table 5-1: Cell Indices

Cell Class

Index

A-3

1.818

A-5

1.818

B-1

4.545

B-3

3.716

B-5

1.244

B-7

1.515

C-3

2.488

C-5

0.606

5.2.3 Removing the Micro-Fuel Cell

The Micro-Fuel cell is located inside the nylon Probe behind the front panel. (See Figure 5-1.) To remove an existing cell:

1.Remove power to the instrument by unplugging the power cord at the power source.

2.Open the front panel door by pressing the release button on the top right corner of the door all the way in and releasing it.

3.Pull up on the nylon Probe, with a slight rocking motion, to release it from the Probe Receptacle.

4.Do Not remove the O-rings unless they are worn and no longer hold the Probe tightly. (If worn, replace them.)

5.When it is free, unscrew the Cap from the nylon Probe. Hold the Probe vertically to prevent dropping the cell out of the probe.

6.Remove the Cell from the Probe, and dispose of it in an environmentally safe manner.

Teledyne Analytical Instruments

5-3

Image 53
Contents Model 3000PA Teledyne Analytical Instruments Model 3000PAInstrument Serial Number Includes the following options IiiPercent Oxygen Analyzer Specific Model Information Installation Model 3000PA Table of Contents IntroductionOperationalTheory OperationAppendix MaintenancePercent Oxygen Analyzer Combustible GAS Usage Warning Overview Main Features of the AnalyzerTypical Applications Percent Oxygen Analyzer IntroductionIntroduction Model 3000PA Model DesignationsModel 3000PA Front Panel Front Panel Operator InterfaceIntroduction Model 3000PA Analog Outputs Recognizing Difference Between LCDRear Panel Equipment Interface Remote Probe RS-232 PortNetwork OptionalMicro-Fuel Cell Sensor Operational Theory IntroductionPercent Oxygen Analyzer Operational Theory Principles of OperationAnatomy of a Micro-Fuel Cell Operational Theory Model 3000PA+ 2H O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsClass A-5 Cell Micro-Fuel Cell ClassClass A-3 Cell Sample System Hydrogen and/or Helium ServiceClass B-1 Cell Class B-3 CellPiping Layout and Flow Diagram for Standard Model Span Zero Sample Electronics and Signal ProcessingBlock Diagram of the Model 3000P Electronics Operational Theory Model 3000PA Mounting the Analyzer Percent Oxygen Analyzer InstallationUnpacking the Analyzer Rear Panel Connections Installation Model 3000PARear Panel of the Model 3000PA Gas ConnectionsElectrical Connections Percent Oxygen Analyzer Installation Teledyne Analytical Instruments Percent Oxygen Analyzer Installation Command Description Remote Probe Connector Parameter SettingTesting the System Installing the Micro-Fuel CellAuto Ranging on Percent Oxygen Analyzer OperationOperation Model 3000PA Using the Data Entry and Function ButtonsHierarchy of Functions and Subfunctions System FunctionSetting the Display Password Protection Percent Oxygen Analyzer Operation Setting up an Auto-CalEntering the Password Installing or Changing the Password Logout Version Screen Cell Failure Span FunctionsSpan Cal Auto Mode SpanningManual Mode Spanning Alarms FunctionInstallation AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Operation Model 3000PA Autoranging AnalysisAnalyze Function Signal OutputVoltage Signal Current Signal Output V dc Output mA dc Range Voltage Current mA Cell Replacement Routine MaintenancePercent Oxygen Analyzer Maintenance Storing and Handling Replacement CellsWhen to Replace a Cell Maintenance Model 3000PRemoving the Micro-Fuel Cell IndexRemoving or Installing a Micro-Fuel Cell Cell Warranty System Self Diagnostic Test Fuse ReplacementMajor Internal Components PowerAnalog PreampRear-Panel Screws CleaningPossible Cause TroubleshootingSolution Maintenance Model 3000P Teledyne Analytical Instruments Percent Oxygen Analyzer Appendix Model 3000PA SpecificationsAppendix Model 3000PA Qty Part Number Description Recommended 2-Year Spare Parts List19-inch Relay Rack Panel Mount Drawing ListSeries Analyzers Calibration GAS Conversions Zero Cal ZeroSettlingMan ENT To Begin