Teledyne 3000PA operating instructions Troubleshooting, Possible Cause, Solution

Page 59

Percent Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

5.7Troubleshooting

Problem:

Erratic readings of the Oxygen concentration as reported by the analyzer.

Possible Cause:

The analyzer may have been calibrated in an inaccurate fashion.

Solution:

Turn the analyzer off, then back on again. Press the System key when prompted by the analyzer "Press System for default Values". This will return the analyzer to its default settings in calibration and zero values. If erratic behavior continues replace the sensor.

Possible Cause:

Atmospheric Oxygen may be diffusing in through the vent and affecting the oxygen level which the sensor sees.

Solution:

Increase flow rate and/or length or vent tubing in order to dilute of mini- mize the diffusion of oxygen from the vent back to the sensor.

Problem:

Inaccurate zero operation (i.e. the user has zeroed the analyzer accidentally on gas much higher than one would normally use for a zero gas).

Solution:

Turn the analyzer off, then back on again. Press the System key when prompted by the analyzer "Press System for default Values". This will return the analyzer to its default settings in calibration and zero values. Now proceed to carefully calibrate and zero the analyzer.

Teledyne Analytical Instruments

5-9

Image 59
Contents Model 3000PA Teledyne Analytical Instruments Model 3000PAInstrument Serial Number Includes the following options IiiPercent Oxygen Analyzer Specific Model Information Operation Model 3000PA Table of Contents IntroductionInstallation OperationalTheoryAppendix MaintenancePercent Oxygen Analyzer Combustible GAS Usage Warning Percent Oxygen Analyzer Introduction Main Features of the AnalyzerOverview Typical ApplicationsIntroduction Model 3000PA Model DesignationsModel 3000PA Front Panel Front Panel Operator InterfaceIntroduction Model 3000PA Analog Outputs Recognizing Difference Between LCDRear Panel Equipment Interface Optional RS-232 PortRemote Probe NetworkPrinciples of Operation Operational Theory IntroductionMicro-Fuel Cell Sensor Percent Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model 3000PA+ 2H O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsClass A-5 Cell Micro-Fuel Cell ClassClass A-3 Cell Class B-3 Cell Hydrogen and/or Helium ServiceSample System Class B-1 CellPiping Layout and Flow Diagram for Standard Model Span Zero Sample Electronics and Signal ProcessingBlock Diagram of the Model 3000P Electronics Operational Theory Model 3000PA Mounting the Analyzer Percent Oxygen Analyzer InstallationUnpacking the Analyzer Rear Panel Connections Installation Model 3000PARear Panel of the Model 3000PA Gas ConnectionsElectrical Connections Percent Oxygen Analyzer Installation Teledyne Analytical Instruments Percent Oxygen Analyzer Installation Command Description Remote Probe Connector Parameter SettingTesting the System Installing the Micro-Fuel CellAuto Ranging on Percent Oxygen Analyzer OperationOperation Model 3000PA Using the Data Entry and Function ButtonsHierarchy of Functions and Subfunctions System FunctionSetting the Display Password Protection Percent Oxygen Analyzer Operation Setting up an Auto-CalEntering the Password Installing or Changing the Password Logout Version Screen Cell Failure Span FunctionsSpan Cal Auto Mode SpanningManual Mode Spanning Alarms FunctionInstallation AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Operation Model 3000PA Autoranging AnalysisAnalyze Function Signal OutputVoltage Signal Current Signal Output V dc Output mA dc Range Voltage Current mA Storing and Handling Replacement Cells Routine MaintenanceCell Replacement Percent Oxygen Analyzer MaintenanceWhen to Replace a Cell Maintenance Model 3000PRemoving the Micro-Fuel Cell IndexRemoving or Installing a Micro-Fuel Cell Cell Warranty System Self Diagnostic Test Fuse ReplacementPreamp PowerMajor Internal Components AnalogRear-Panel Screws CleaningPossible Cause TroubleshootingSolution Maintenance Model 3000P Teledyne Analytical Instruments Percent Oxygen Analyzer Appendix Model 3000PA SpecificationsAppendix Model 3000PA Qty Part Number Description Recommended 2-Year Spare Parts List19-inch Relay Rack Panel Mount Drawing ListSeries Analyzers Calibration GAS Conversions Zero Cal ZeroSettlingMan ENT To Begin