Teledyne 3000PA operating instructions AL-1 AL-2 Choose Alarm

Page 46

4 Operation

Model 3000PA

 

 

 

 

 

 

 

 

 

The defeat function can also be used to reset a latched alarm. (See procedures, below.)

If you are using password protection, you will need to enter your password to access the alarm functions. Follow the instructions in section

4.3.3to enter your password. Once you have clearance to proceed, enter the Alarm function.

Press the Alarm button on the front panel to enter the Alarm function. Make sure that AL–1is blinking.

AL—1 AL—2

Choose Alarm

Set up alarm 1 by moving the blinking over to AL–1using the < > arrow keys. Then press Enter to move to the next screen.

AL—1 1.00 % HI

Dft—N Fs—N Ltch—N

Five parameters can be changed on this screen:

• Value of the alarm setpoint, AL–1 #### % (oxygen)

Out-of-range direction, HI or LO

Defeated? Dft–Y/N(Yes/No)

Failsafe? Fs–Y/N(Yes/No)

Latching? Ltch–Y/N(Yes/No).

To define the setpoint, use the < > arrow keys to move the

blinking over to AL–1 ####. Then use the ∆∇ arrow keys to change the number. Holding down the key speeds up the incrementing or decrementing. (Remember, the setpoint units are percent-of-oxygen.)

To set the other parameters use the < > arrow keys to move the

blinking over to the desired parameter. Then use the ∆∇ arrow keys to change the parameter.

Once the parameters for alarm 1 have been set, press Alarms again, and repeat this procedure for alarm 2 (AL–2).

To reset a latched alarm, go to Dft– and then press either two times or two times. (Toggle it to Y and then back to N.)

–OR

Go to Ltch– and then press either two times or two times. (Toggle it to N and back to Y.)

4-14

Teledyne Analytical Instruments

Image 46
Contents Model 3000PA Model 3000PA Teledyne Analytical InstrumentsPercent Oxygen Analyzer Specific Model Information IiiInstrument Serial Number Includes the following options OperationalTheory Model 3000PA Table of Contents IntroductionInstallation OperationPercent Oxygen Analyzer MaintenanceAppendix Combustible GAS Usage Warning Typical Applications Main Features of the AnalyzerOverview Percent Oxygen Analyzer IntroductionModel Designations Introduction Model 3000PAFront Panel Operator Interface Model 3000PA Front PanelIntroduction Model 3000PA Rear Panel Equipment Interface Recognizing Difference Between LCDAnalog Outputs Network RS-232 PortRemote Probe OptionalPercent Oxygen Analyzer Operational Theory Operational Theory IntroductionMicro-Fuel Cell Sensor Principles of OperationOperational Theory Model 3000PA Anatomy of a Micro-Fuel CellElectrochemical Reactions + 2H O + 4e → 4OHCalibration Characteristics Effect of PressureClass A-3 Cell Micro-Fuel Cell ClassClass A-5 Cell Class B-1 Cell Hydrogen and/or Helium ServiceSample System Class B-3 CellPiping Layout and Flow Diagram for Standard Model Electronics and Signal Processing Span Zero SampleBlock Diagram of the Model 3000P Electronics Operational Theory Model 3000PA Unpacking the Analyzer Percent Oxygen Analyzer InstallationMounting the Analyzer Installation Model 3000PA Rear Panel ConnectionsGas Connections Rear Panel of the Model 3000PAElectrical Connections Percent Oxygen Analyzer Installation Teledyne Analytical Instruments Percent Oxygen Analyzer Installation Command Description Parameter Setting Remote Probe ConnectorInstalling the Micro-Fuel Cell Testing the SystemPercent Oxygen Analyzer Operation Auto Ranging onUsing the Data Entry and Function Buttons Operation Model 3000PASystem Function Hierarchy of Functions and SubfunctionsSetting the Display Percent Oxygen Analyzer Operation Setting up an Auto-Cal Password ProtectionEntering the Password Installing or Changing the Password Logout Version Screen Span Functions Cell FailureAuto Mode Spanning Span CalAlarms Function Manual Mode SpanningInstallation AL-1 AL-2 Choose Alarm Setting the Analog Output Ranges Range FunctionOperation Model 3000PA Autoranging Analysis Fixed Range AnalysisVoltage Signal Current Signal Output V dc Output mA dc Signal OutputAnalyze Function Range Voltage Current mA Percent Oxygen Analyzer Maintenance Routine MaintenanceCell Replacement Storing and Handling Replacement CellsMaintenance Model 3000P When to Replace a CellIndex Removing the Micro-Fuel CellRemoving or Installing a Micro-Fuel Cell Cell Warranty Fuse Replacement System Self Diagnostic TestAnalog PowerMajor Internal Components PreampCleaning Rear-Panel ScrewsSolution TroubleshootingPossible Cause Maintenance Model 3000P Teledyne Analytical Instruments Model 3000PA Specifications Percent Oxygen Analyzer AppendixAppendix Model 3000PA Recommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountSeries Analyzers Calibration GAS Conversions Zero Cal ZeroSettlingMan ENT To Begin