4 Operation

Model 3000PA

 

 

 

 

 

 

 

 

 

of the displayed value of the oxygen concentration (in ppm) for three min- utes. Then the instrument automatically returns to the analyze mode.

4.4.2.2Manual Mode Spanning

Press Span to start the Span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually.

Span: Settling:MAN

<ENT> For Next

Use the ∆∇ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen.

Span Val: 20.90 <ENT>Span <UP>Mod #

Press (<UP>) to permit modification (Mod #) of span value.

Use the arrow keys to enter the oxygen concentration of the span gas you are using (20.90 if you are using air). The < > arrows chose the digit, and the ∆∇ arrows choose the value of the digit.

Press Enter to enter the span value into the system and begin the span calibration.

Once the span has begun, the microprocessor samples the output at a predetermined rate. It calculates the difference between successive samplings and displays this difference as Slope on the screen. It takes several seconds for the first Slope value to display. Slope indicates rate of change of the Span reading. It is a sensitive indicator of stability.

####% Span

Slope=#### ppm/s

When the Span value displayed on the screen is sufficiently stable, press Enter. (Generally, when the Span reading changes by 1 % or less of the full scale of the range being calibrated for a period of five minutes it is sufficiently stable.) Once Enter is pressed, the Span reading changes to the correct value. The instrument then automatically enters the Analyze func- tion.

4.5The Alarms Function

The Model 3000PA is equipped with 2 fully adjustable concentration alarms and a system failure alarm. Each alarm has a relay with a set of form

4-12

Teledyne Analytical Instruments

Image 44
Contents Model 3000PA Model 3000PA Teledyne Analytical InstrumentsInstrument Serial Number Includes the following options IiiPercent Oxygen Analyzer Specific Model Information Model 3000PA Table of Contents Introduction InstallationOperationalTheory OperationAppendix MaintenancePercent Oxygen Analyzer Combustible GAS Usage Warning Main Features of the Analyzer OverviewTypical Applications Percent Oxygen Analyzer IntroductionModel Designations Introduction Model 3000PAFront Panel Operator Interface Model 3000PA Front PanelIntroduction Model 3000PA Analog Outputs Recognizing Difference Between LCDRear Panel Equipment Interface RS-232 Port Remote ProbeNetwork OptionalOperational Theory Introduction Micro-Fuel Cell SensorPercent Oxygen Analyzer Operational Theory Principles of OperationOperational Theory Model 3000PA Anatomy of a Micro-Fuel CellElectrochemical Reactions + 2H O + 4e → 4OHCalibration Characteristics Effect of PressureClass A-5 Cell Micro-Fuel Cell ClassClass A-3 Cell Hydrogen and/or Helium Service Sample SystemClass B-1 Cell Class B-3 CellPiping Layout and Flow Diagram for Standard Model Electronics and Signal Processing Span Zero SampleBlock Diagram of the Model 3000P Electronics Operational Theory Model 3000PA Mounting the Analyzer Percent Oxygen Analyzer InstallationUnpacking the Analyzer Installation Model 3000PA Rear Panel ConnectionsGas Connections Rear Panel of the Model 3000PAElectrical Connections Percent Oxygen Analyzer Installation Teledyne Analytical Instruments Percent Oxygen Analyzer Installation Command Description Parameter Setting Remote Probe ConnectorInstalling the Micro-Fuel Cell Testing the SystemPercent Oxygen Analyzer Operation Auto Ranging onUsing the Data Entry and Function Buttons Operation Model 3000PASystem Function Hierarchy of Functions and SubfunctionsSetting the Display Percent Oxygen Analyzer Operation Setting up an Auto-Cal Password ProtectionEntering the Password Installing or Changing the Password Logout Version Screen Span Functions Cell FailureAuto Mode Spanning Span CalAlarms Function Manual Mode SpanningInstallation AL-1 AL-2 Choose Alarm Setting the Analog Output Ranges Range FunctionOperation Model 3000PA Autoranging Analysis Fixed Range AnalysisAnalyze Function Signal OutputVoltage Signal Current Signal Output V dc Output mA dc Range Voltage Current mA Routine Maintenance Cell ReplacementPercent Oxygen Analyzer Maintenance Storing and Handling Replacement CellsMaintenance Model 3000P When to Replace a CellIndex Removing the Micro-Fuel CellRemoving or Installing a Micro-Fuel Cell Cell Warranty Fuse Replacement System Self Diagnostic TestPower Major Internal ComponentsAnalog PreampCleaning Rear-Panel ScrewsPossible Cause TroubleshootingSolution Maintenance Model 3000P Teledyne Analytical Instruments Model 3000PA Specifications Percent Oxygen Analyzer AppendixAppendix Model 3000PA Recommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountSeries Analyzers Calibration GAS Conversions Zero Cal ZeroSettlingMan ENT To Begin