Teledyne 3000PA operating instructions ZeroSettlingMan ENT To Begin

Page 69

Percent Oxygen Analyzer

Appendix

 

 

 

 

 

 

 

 

 

to toggle between AUTO and MAN zero settling. Stop when MAN appears, blinking, on the display.

Zero:Settling:Man

<ENT> To Begin

Press Enter to begin the zero calibration. After a few seconds the first of five zeroing screens appears. The number in the upper left hand corner is the first-stage zero offset. The microprocessor samples the output at a prede-

termined rate. It calculates the differences between successive samplings and displays the rate of change as Slope= a value in parts per million per second

(ppm/s).

####% Zero

Slope=#### ppm/s

NOTE: It takes several seconds for the true Slope value to display. Wait about 10 seconds. Then, wait until Slope is sufficiently close to zero before pressing Enter to finish zeroing .

Generally, you have a good zero when Slope is less than 0.05 ppm/s for about 30 seconds. When Slope is close enough to zero, press Enter. In a few seconds, the screen will update.

Once span settling completes, the information is stored in the microprocessor, and the instrument automatically returns to the Analyze mode.

NOTE: The MSDS on this material is available upon request through the Teledyne Environmental, Health and Safety Coordinator. Contact at (626) 934-1592

Teledyne Analytical Instruments

A-9

Image 69
Contents Model 3000PA Teledyne Analytical Instruments Model 3000PAIii Percent Oxygen Analyzer Specific Model InformationInstrument Serial Number Includes the following options Installation Model 3000PA Table of Contents IntroductionOperationalTheory OperationMaintenance Percent Oxygen AnalyzerAppendix Combustible GAS Usage Warning Overview Main Features of the AnalyzerTypical Applications Percent Oxygen Analyzer IntroductionIntroduction Model 3000PA Model DesignationsModel 3000PA Front Panel Front Panel Operator InterfaceIntroduction Model 3000PA Recognizing Difference Between LCD Rear Panel Equipment InterfaceAnalog Outputs Remote Probe RS-232 PortNetwork OptionalMicro-Fuel Cell Sensor Operational Theory IntroductionPercent Oxygen Analyzer Operational Theory Principles of OperationAnatomy of a Micro-Fuel Cell Operational Theory Model 3000PA+ 2H O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsMicro-Fuel Cell Class Class A-3 CellClass A-5 Cell Sample System Hydrogen and/or Helium ServiceClass B-1 Cell Class B-3 CellPiping Layout and Flow Diagram for Standard Model Span Zero Sample Electronics and Signal ProcessingBlock Diagram of the Model 3000P Electronics Operational Theory Model 3000PA Percent Oxygen Analyzer Installation Unpacking the AnalyzerMounting the Analyzer Rear Panel Connections Installation Model 3000PARear Panel of the Model 3000PA Gas ConnectionsElectrical Connections Percent Oxygen Analyzer Installation Teledyne Analytical Instruments Percent Oxygen Analyzer Installation Command Description Remote Probe Connector Parameter SettingTesting the System Installing the Micro-Fuel CellAuto Ranging on Percent Oxygen Analyzer OperationOperation Model 3000PA Using the Data Entry and Function ButtonsHierarchy of Functions and Subfunctions System FunctionSetting the Display Password Protection Percent Oxygen Analyzer Operation Setting up an Auto-CalEntering the Password Installing or Changing the Password Logout Version Screen Cell Failure Span FunctionsSpan Cal Auto Mode SpanningManual Mode Spanning Alarms FunctionInstallation AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Operation Model 3000PA Autoranging AnalysisSignal Output Voltage Signal Current Signal Output V dc Output mA dcAnalyze Function Range Voltage Current mA Cell Replacement Routine MaintenancePercent Oxygen Analyzer Maintenance Storing and Handling Replacement CellsWhen to Replace a Cell Maintenance Model 3000PRemoving the Micro-Fuel Cell IndexRemoving or Installing a Micro-Fuel Cell Cell Warranty System Self Diagnostic Test Fuse ReplacementMajor Internal Components PowerAnalog PreampRear-Panel Screws CleaningTroubleshooting SolutionPossible Cause Maintenance Model 3000P Teledyne Analytical Instruments Percent Oxygen Analyzer Appendix Model 3000PA SpecificationsAppendix Model 3000PA Qty Part Number Description Recommended 2-Year Spare Parts List19-inch Relay Rack Panel Mount Drawing ListSeries Analyzers Calibration GAS Conversions Zero Cal ZeroSettlingMan ENT To Begin