Percent Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

5.2.4 Installing a New Micro-Fuel Cell

CAUTION: Do not touch the sensing surface of the cell. It is covered with a delicate Teflon membrane that can leak if punctured. The sensor must be replaced if the membrane is damaged.

1.Place the Cell in the Probe with the sensing surface facing outward (toward the screen in the Cap).

2.Screw the Probe Cap onto the Probe until it stops.

3.With the O-rings in place, push the assembled Probe down into the Cell Holder—Cap Down—with a slight rocking motion until it is seated on the bottom of the holder. This forces the holder into position and forms a gas-tight seal.

5.2.5 Cell Warranty

The Class B-1 Micro-Fuel cell is standard in the Model 3000PA. This cell is warranted for 6 months from the date of shipment. Check the Spe- cific Model Information, and note any Addendum that might be attached to the front of this manual for special information applying to the Cell in your instrument.

The warranty period for spare cells begins on the date of shipment. Do not purchase more than one spare cell per instrument. Do not stockpile spare cells.

The B-1 cell is not designed for applications where CO2 is a major

component in the sample, however slight amounts will not adversely effect the cell performance. Consult TAI for available options for either intermittent or continuous CO2 exposure.

If a cell was working satisfactorily, but ceases to function before the warranty period expires, the customer will receive credit toward the pur- chase of a new cell.

If you have a warranty claim, you must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty workmanship or material, the cell will be replaced at no cost to you.

NOTE: Evidence of damage due to tampering or mishandling will render the cell warranty null and void.

Teledyne Analytical Instruments

5-5

Image 55
Contents Model 3000PA Teledyne Analytical Instruments Model 3000PAPercent Oxygen Analyzer Specific Model Information IiiInstrument Serial Number Includes the following options Operation Model 3000PA Table of Contents IntroductionInstallation OperationalTheoryPercent Oxygen Analyzer MaintenanceAppendix Combustible GAS Usage Warning Percent Oxygen Analyzer Introduction Main Features of the AnalyzerOverview Typical ApplicationsIntroduction Model 3000PA Model DesignationsModel 3000PA Front Panel Front Panel Operator InterfaceIntroduction Model 3000PA Rear Panel Equipment Interface Recognizing Difference Between LCDAnalog Outputs Optional RS-232 PortRemote Probe NetworkPrinciples of Operation Operational Theory IntroductionMicro-Fuel Cell Sensor Percent Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model 3000PA+ 2H O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsClass A-3 Cell Micro-Fuel Cell ClassClass A-5 Cell Class B-3 Cell Hydrogen and/or Helium ServiceSample System Class B-1 CellPiping Layout and Flow Diagram for Standard Model Span Zero Sample Electronics and Signal ProcessingBlock Diagram of the Model 3000P Electronics Operational Theory Model 3000PA Unpacking the Analyzer Percent Oxygen Analyzer InstallationMounting the Analyzer Rear Panel Connections Installation Model 3000PARear Panel of the Model 3000PA Gas ConnectionsElectrical Connections Percent Oxygen Analyzer Installation Teledyne Analytical Instruments Percent Oxygen Analyzer Installation Command Description Remote Probe Connector Parameter SettingTesting the System Installing the Micro-Fuel CellAuto Ranging on Percent Oxygen Analyzer OperationOperation Model 3000PA Using the Data Entry and Function ButtonsHierarchy of Functions and Subfunctions System FunctionSetting the Display Password Protection Percent Oxygen Analyzer Operation Setting up an Auto-CalEntering the Password Installing or Changing the Password Logout Version Screen Cell Failure Span FunctionsSpan Cal Auto Mode SpanningManual Mode Spanning Alarms FunctionInstallation AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Operation Model 3000PA Autoranging AnalysisVoltage Signal Current Signal Output V dc Output mA dc Signal OutputAnalyze Function Range Voltage Current mA Storing and Handling Replacement Cells Routine MaintenanceCell Replacement Percent Oxygen Analyzer Maintenance When to Replace a Cell Maintenance Model 3000PRemoving the Micro-Fuel Cell IndexRemoving or Installing a Micro-Fuel Cell Cell Warranty System Self Diagnostic Test Fuse ReplacementPreamp PowerMajor Internal Components AnalogRear-Panel Screws CleaningSolution TroubleshootingPossible Cause Maintenance Model 3000P Teledyne Analytical Instruments Percent Oxygen Analyzer Appendix Model 3000PA SpecificationsAppendix Model 3000PA Qty Part Number Description Recommended 2-Year Spare Parts List19-inch Relay Rack Panel Mount Drawing ListSeries Analyzers Calibration GAS Conversions Zero Cal ZeroSettlingMan ENT To Begin