Percent Oxygen Analyzer | Maintenance 5 | |
|
|
|
|
|
|
|
|
|
5.2.4 Installing a New
CAUTION: Do not touch the sensing surface of the cell. It is covered with a delicate Teflon membrane that can leak if punctured. The sensor must be replaced if the membrane is damaged.
1.Place the Cell in the Probe with the sensing surface facing outward (toward the screen in the Cap).
2.Screw the Probe Cap onto the Probe until it stops.
3.With the
5.2.5 Cell Warranty
The Class
The warranty period for spare cells begins on the date of shipment. Do not purchase more than one spare cell per instrument. Do not stockpile spare cells.
The
component in the sample, however slight amounts will not adversely effect the cell performance. Consult TAI for available options for either intermittent or continuous CO2 exposure.
If a cell was working satisfactorily, but ceases to function before the warranty period expires, the customer will receive credit toward the pur- chase of a new cell.
If you have a warranty claim, you must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty workmanship or material, the cell will be replaced at no cost to you.
NOTE: Evidence of damage due to tampering or mishandling will render the cell warranty null and void.