Teledyne 3000PA operating instructions Calibration GAS

Page 66

Appendix

Model 3000PA

 

 

 

 

 

 

 

 

 

RESTRICTOR KIT

The current revision of the 3000 series analyzers are supplied with a kit containing two restrictors and a union which are user installed. These parts supplied to give the end user more flexibility when installing the analyzer. The restrictor kit is suitable for high and low positive pressure applications as well as vacuum service ( atmospheric pressure sample) applications ( see manual for installation instructions). The standard restrictor ( BLUE DOT ) is recommended for pressures between 5 PSIG and 50 PSIG. For positive low pressure application ( 5 psig or less ) the un-marked restrictor is better suited . For none pressurized sample applications the marked restrictor should be used and configured for vacuum service. Note: for extremely low positive pressure applications ( less then 2 psig) the vacuum service configuration should provide higher performance ( higher flow rates). For vacuum service the end user must supply a vacuum pump and a by-pass valve for the pump. A vacuum level of 5 - 10 inches of mercury should provide the optimum flow rate. CAUTION: flow restrictors have very small orifices and may be plugged by small particles ( .005” dia or larger) A sample filter must be included in the sample line prior to the restrictor! ( a 60 micron filter is recommended)

3000PA EXAMPLES:

Example 1, with a incoming pressure of 10 psig the std restrictor (blue dot) will provide a flow rate of .76 SLPM. Up-stream of the restrictor the sample line pressure will be 10 psig, while down stream ( including the cell) the pressure will be at atmospheric pressure.( analyzer vented to atmospheric pressure) Note, all other pressure drops in the sample path are insignificant at these flow rates. This insures that the cell operates at atmospheric pressure. At very high flow rates ( off scale of flow-meter), pressure drops other than the restriction device could become significant , and result in pressurizing the cell.

Example 2, A 3000PA is configured for vacuum service as follows. The un-marked restrictor is placed in the sample vent port. The down stream end of the restrictor is then connected to a vacuum pump and by-pass valve. The by- pass valve is adjusted to provide a flow rate of 1 SLPM. The sample pressure between the pump and the restrictor will be approximately -7 inches of mercury, while the pressure in the balance of the sample system including the cell will be approximately at atmospheric pressure. ( provided the sample flow into the analyzer is not blocked.)

BY-PASS:

To improve the system response, a by-pass can be added to increase the sample flow rate to the analyzer by a factor of ten. A by-pass provides a sample flow path around the analyzer of 2 - 18 SCFH. typically.

CALIBRATION GAS:

3000 series analyzer requirements for units with Auto-Cal options. The

A-6

Teledyne Analytical Instruments

Image 66
Contents Model 3000PA Model 3000PA Teledyne Analytical InstrumentsIii Percent Oxygen Analyzer Specific Model InformationInstrument Serial Number Includes the following options OperationalTheory Model 3000PA Table of Contents IntroductionInstallation OperationMaintenance Percent Oxygen AnalyzerAppendix Combustible GAS Usage Warning Typical Applications Main Features of the AnalyzerOverview Percent Oxygen Analyzer IntroductionModel Designations Introduction Model 3000PAFront Panel Operator Interface Model 3000PA Front PanelIntroduction Model 3000PA Recognizing Difference Between LCD Rear Panel Equipment InterfaceAnalog Outputs Network RS-232 PortRemote Probe OptionalPercent Oxygen Analyzer Operational Theory Operational Theory IntroductionMicro-Fuel Cell Sensor Principles of OperationOperational Theory Model 3000PA Anatomy of a Micro-Fuel CellElectrochemical Reactions + 2H O + 4e → 4OHCalibration Characteristics Effect of PressureMicro-Fuel Cell Class Class A-3 CellClass A-5 Cell Class B-1 Cell Hydrogen and/or Helium ServiceSample System Class B-3 CellPiping Layout and Flow Diagram for Standard Model Electronics and Signal Processing Span Zero SampleBlock Diagram of the Model 3000P Electronics Operational Theory Model 3000PA Percent Oxygen Analyzer Installation Unpacking the AnalyzerMounting the Analyzer Installation Model 3000PA Rear Panel ConnectionsGas Connections Rear Panel of the Model 3000PAElectrical Connections Percent Oxygen Analyzer Installation Teledyne Analytical Instruments Percent Oxygen Analyzer Installation Command Description Parameter Setting Remote Probe ConnectorInstalling the Micro-Fuel Cell Testing the SystemPercent Oxygen Analyzer Operation Auto Ranging onUsing the Data Entry and Function Buttons Operation Model 3000PASystem Function Hierarchy of Functions and SubfunctionsSetting the Display Percent Oxygen Analyzer Operation Setting up an Auto-Cal Password ProtectionEntering the Password Installing or Changing the Password Logout Version Screen Span Functions Cell FailureAuto Mode Spanning Span CalAlarms Function Manual Mode SpanningInstallation AL-1 AL-2 Choose Alarm Setting the Analog Output Ranges Range FunctionOperation Model 3000PA Autoranging Analysis Fixed Range AnalysisSignal Output Voltage Signal Current Signal Output V dc Output mA dcAnalyze Function Range Voltage Current mA Percent Oxygen Analyzer Maintenance Routine MaintenanceCell Replacement Storing and Handling Replacement CellsMaintenance Model 3000P When to Replace a CellIndex Removing the Micro-Fuel CellRemoving or Installing a Micro-Fuel Cell Cell Warranty Fuse Replacement System Self Diagnostic TestAnalog PowerMajor Internal Components PreampCleaning Rear-Panel ScrewsTroubleshooting SolutionPossible Cause Maintenance Model 3000P Teledyne Analytical Instruments Model 3000PA Specifications Percent Oxygen Analyzer AppendixAppendix Model 3000PA Recommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountSeries Analyzers Calibration GAS Conversions Zero Cal ZeroSettlingMan ENT To Begin