4 Operation

Model 3000PA

 

 

 

 

 

 

 

 

 

4.6.2 Autoranging Analysis

Set your analysis ranges as in 4.6.1, above. Leave Mode in Auto, or use the arrow buttons to change back to Auto.

When operating in autoranging, if the oxygen concentration in your sample goes ABOVE your HIGHEST range setting, the analyzer will go into the special 25 % cal range.

However, if one of your range settings is below 0-25 % and another is set above 0-25 %, the special 0-25 % Air Cal range will NOT activate as the oxygen level goes through 25 %. Nevertheless, if the oxygen concentration in your sample goes ABOVE your HIGHEST range setting, the analyzer will THEN drop back down into the special 25 % cal range.

Once the oxygen concentration drops back down into your highest range setting, the analyzer will automatically switch back to that range.

CAUTION: While the analyzer is in the Air Cal range, the oxygen reading cannot go over 25 %, even if the oxygen concentration is higher than 25 %.

4.6.3 Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing the analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, enter the range function by pressing the Range button on the front panel.

Use the < > arrow keys to move the blinking over AUTO.

Use the ∆∇ arrow keys to switch from AUTO to FX/LO, FX/MED, or FX/HI to set the instrument on the desired fixed range (low, medium, or high).

L—1.00 M—5.00

H—10.00 Mode—FX/LO

or

L—1.00 M—5.00

H—10.00 Mode—FX/MED

or

L—1.00 M—5.00

H—10.00 Mode—FX/HI

4-16

Teledyne Analytical Instruments

Image 48
Contents Model 3000PA Model 3000PA Teledyne Analytical InstrumentsIii Percent Oxygen Analyzer Specific Model InformationInstrument Serial Number Includes the following options Model 3000PA Table of Contents Introduction InstallationOperationalTheory OperationMaintenance Percent Oxygen AnalyzerAppendix Combustible GAS Usage Warning Main Features of the Analyzer OverviewTypical Applications Percent Oxygen Analyzer IntroductionModel Designations Introduction Model 3000PAFront Panel Operator Interface Model 3000PA Front PanelIntroduction Model 3000PA Recognizing Difference Between LCD Rear Panel Equipment InterfaceAnalog Outputs RS-232 Port Remote ProbeNetwork OptionalOperational Theory Introduction Micro-Fuel Cell SensorPercent Oxygen Analyzer Operational Theory Principles of OperationOperational Theory Model 3000PA Anatomy of a Micro-Fuel CellElectrochemical Reactions + 2H O + 4e → 4OHCalibration Characteristics Effect of PressureMicro-Fuel Cell Class Class A-3 CellClass A-5 Cell Hydrogen and/or Helium Service Sample SystemClass B-1 Cell Class B-3 CellPiping Layout and Flow Diagram for Standard Model Electronics and Signal Processing Span Zero SampleBlock Diagram of the Model 3000P Electronics Operational Theory Model 3000PA Percent Oxygen Analyzer Installation Unpacking the AnalyzerMounting the Analyzer Installation Model 3000PA Rear Panel ConnectionsGas Connections Rear Panel of the Model 3000PAElectrical Connections Percent Oxygen Analyzer Installation Teledyne Analytical Instruments Percent Oxygen Analyzer Installation Command Description Parameter Setting Remote Probe ConnectorInstalling the Micro-Fuel Cell Testing the SystemPercent Oxygen Analyzer Operation Auto Ranging onUsing the Data Entry and Function Buttons Operation Model 3000PASystem Function Hierarchy of Functions and SubfunctionsSetting the Display Percent Oxygen Analyzer Operation Setting up an Auto-Cal Password ProtectionEntering the Password Installing or Changing the Password Logout Version Screen Span Functions Cell FailureAuto Mode Spanning Span CalAlarms Function Manual Mode SpanningInstallation AL-1 AL-2 Choose Alarm Setting the Analog Output Ranges Range FunctionOperation Model 3000PA Autoranging Analysis Fixed Range AnalysisSignal Output Voltage Signal Current Signal Output V dc Output mA dcAnalyze Function Range Voltage Current mA Routine Maintenance Cell ReplacementPercent Oxygen Analyzer Maintenance Storing and Handling Replacement CellsMaintenance Model 3000P When to Replace a CellIndex Removing the Micro-Fuel CellRemoving or Installing a Micro-Fuel Cell Cell Warranty Fuse Replacement System Self Diagnostic TestPower Major Internal ComponentsAnalog PreampCleaning Rear-Panel ScrewsTroubleshooting SolutionPossible Cause Maintenance Model 3000P Teledyne Analytical Instruments Model 3000PA Specifications Percent Oxygen Analyzer AppendixAppendix Model 3000PA Recommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountSeries Analyzers Calibration GAS Conversions Zero Cal ZeroSettlingMan ENT To Begin