Teledyne operating instructions Model 3000PA Specifications, Percent Oxygen Analyzer Appendix

Page 61

Percent Oxygen Analyzer

Appendix

 

 

 

 

 

 

 

 

 

Appendix

A-1 Model 3000PA Specifications

Packaging:

General Purpose

 

• Flush panel mount (Standard).

 

• Rack mount — Relay rack mounted to

 

contain either one or two instruments in one

 

19" relay rack mountable plate (Optional).

Sensor:

Class B-1 Micro-Fuel Cell (standard). Others

 

available.

Cell Block:

Nylon.

90 % Response Time:

10 seconds at 25 °C (77 °F).

Ranges:

Three user definable ranges from 0-1 % to

 

0-100 %, plus air calibration range of 0-25 %.

Alarms:

One system-failure alarm contact to detect

 

power failure.

 

Two adjustable concentration threshold alarms

 

with fully programmable setpoints.

Displays:

2 line by 20 character, alphanumeric, VFD

 

screen.

 

One 5 digit LED display.

Digital Interface:

Full duplex RS-232 communications port.

Power: Universal power supply 85-250 V ac, at

 

47-63 Hz.

Operating Temperature:

0-50 °C (32-122 °F).

Teledyne Analytical Instruments

A-1

Image 61
Contents Model 3000PA Teledyne Analytical Instruments Model 3000PAPercent Oxygen Analyzer Specific Model Information IiiInstrument Serial Number Includes the following options Installation Model 3000PA Table of Contents IntroductionOperationalTheory OperationPercent Oxygen Analyzer MaintenanceAppendix Combustible GAS Usage Warning Overview Main Features of the AnalyzerTypical Applications Percent Oxygen Analyzer IntroductionIntroduction Model 3000PA Model DesignationsModel 3000PA Front Panel Front Panel Operator InterfaceIntroduction Model 3000PA Rear Panel Equipment Interface Recognizing Difference Between LCDAnalog Outputs Remote Probe RS-232 PortNetwork OptionalMicro-Fuel Cell Sensor Operational Theory IntroductionPercent Oxygen Analyzer Operational Theory Principles of OperationAnatomy of a Micro-Fuel Cell Operational Theory Model 3000PA+ 2H O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsClass A-3 Cell Micro-Fuel Cell ClassClass A-5 Cell Sample System Hydrogen and/or Helium ServiceClass B-1 Cell Class B-3 CellPiping Layout and Flow Diagram for Standard Model Span Zero Sample Electronics and Signal ProcessingBlock Diagram of the Model 3000P Electronics Operational Theory Model 3000PA Unpacking the Analyzer Percent Oxygen Analyzer InstallationMounting the Analyzer Rear Panel Connections Installation Model 3000PARear Panel of the Model 3000PA Gas ConnectionsElectrical Connections Percent Oxygen Analyzer Installation Teledyne Analytical Instruments Percent Oxygen Analyzer Installation Command Description Remote Probe Connector Parameter SettingTesting the System Installing the Micro-Fuel CellAuto Ranging on Percent Oxygen Analyzer OperationOperation Model 3000PA Using the Data Entry and Function ButtonsHierarchy of Functions and Subfunctions System FunctionSetting the Display Password Protection Percent Oxygen Analyzer Operation Setting up an Auto-CalEntering the Password Installing or Changing the Password Logout Version Screen Cell Failure Span FunctionsSpan Cal Auto Mode SpanningManual Mode Spanning Alarms FunctionInstallation AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Operation Model 3000PA Autoranging AnalysisVoltage Signal Current Signal Output V dc Output mA dc Signal OutputAnalyze Function Range Voltage Current mA Cell Replacement Routine MaintenancePercent Oxygen Analyzer Maintenance Storing and Handling Replacement CellsWhen to Replace a Cell Maintenance Model 3000PRemoving the Micro-Fuel Cell IndexRemoving or Installing a Micro-Fuel Cell Cell Warranty System Self Diagnostic Test Fuse ReplacementMajor Internal Components PowerAnalog PreampRear-Panel Screws CleaningSolution TroubleshootingPossible Cause Maintenance Model 3000P Teledyne Analytical Instruments Percent Oxygen Analyzer Appendix Model 3000PA SpecificationsAppendix Model 3000PA Qty Part Number Description Recommended 2-Year Spare Parts List19-inch Relay Rack Panel Mount Drawing ListSeries Analyzers Calibration GAS Conversions Zero Cal ZeroSettlingMan ENT To Begin