SAFETY INSPECTION/ MAINTENANCE
PROCEDURE 2 | SAFETY INSPECTION/MAINTENANCE |
TROUBLESHOOTING GUIDE
SYMPTOM | PROBABLE CAUSE | SOLUTION | |
|
|
| |
LIMITED DRIVING | Batteries not charged long enough. | Charge batteries overnight or | |
DISTANCE. |
| ensure 8 hours of charge between | |
|
| use times (PROCEDURE 4). | |
| Batteries weak, won't hold charge. | Replace Batteries | |
|
| (PROCEDURE 4). | |
|
|
| |
BATTERIES NOT | Charger not working. | Replace Charger. | |
CHARGING. | Battery Connections loose. | Check all connections. Secure | |
| |||
|
| connections (PROCEDURE 4). | |
| NO current at wall outlet. | Switch to another wall outlet. | |
| BAD connection on Charger, | Replace Charger or internal | |
| Charger Cable, Plug or Internal | repairs required. Contact Dealer | |
| Wiring problem. | or Invacare. | |
|
|
| |
BATTERIES DRAW | Battery Failure. | Replace Batteries | |
EXCESSIVE CURRENT |
| (PROCEDURE 4). | |
WHEN CHARGING. |
|
| |
|
|
| |
CHARGE INDICATOR | Batteries weak, won't hold charge. | Replace Batteries | |
SHOWS LOW CHARGE |
| (PROCEDURE 4). | |
LEVEL IMMEDIATELY | Charger not operating. | Replace Charger. | |
AFTER CHARGING. | |||
|
| ||
|
|
| |
BATTERY INDICATOR | Have charger checked. | Service or replace charger. | |
FLASHES THE CHARGE |
| Contact Dealer or Invacare. | |
LEVEL IS LOW - TOO | Weak batteries. | Replace batteries (PROCEDURE | |
SOON AFTER BEING |
| 4). | |
RECHARGED. |
|
| |
|
|
| |
SCOOTER WILL NOT | Engage | ||
DRIVE. |
| (procedure 3) | |
| Batteries require charging. | Charge batteries (PROCEDURE 4). | |
| Key not fully plugged in. | Ensure key is inserted fully into | |
|
| power plug jack. | |
| Charger plugged in. | Unplug charger from scooter. | |
| Circuit breaker tripped. | Reset breaker. If breaker trips | |
|
| again, it may indicate need for | |
|
| internal repairs. Contact Dealer | |
|
| or Invacare. | |
| Soft stop is engaged.If key is turned | With key in the "OFF" position, | |
| "OFF" while driving, the scooter | wait 30 seconds and turn key to | |
| artificially holds power for 1.5 seconds | "ON" position. | |
| slowly decelerating unit to a stop. |
| |
|
|
| |
MOTOR “CHATTERS” | Fuse is Blown. | Replace Fuse (PROCEDURE 3). | |
OR RUNS | Electrical malfunction. | Contact Dealer or Invacare. | |
IRREGULARLY. |
|
| |
|
|
|
Invacare Scooters | 18 | Part No. 1090132 Rev F |