3M™ Rapid Detection Reader

Working with Results and Logs

Manually Transferring / Printing Results or Logs

Results (assay, patient or IQC) that have been selected on the Results screen can be transferred to a printer, USB flash drive or to a network. It is important to note that the user should manually delete the results from the Reader after manual transfer or they will be transferred again, unless a specific range of results is selected when the next manual transfer is performed.

Log events (All, Failed Assays or Failed LQC) that have been selected on the Events screen can be transferred to a USB flash drive or to a network. Only Failed Assays and Failed LQC can be printed from the CM.

NOTE: A single USB flash drive can be used to transfer results from all Readers as the file names start with their respective control module serial numbers. The USB flash drive must be preconfigured with 4 subdirectories (folders) named: patient, lqc, iqc and log, for this use.

CAUTION.

Do not remove the USB flash drive from the control module while data is being transferred.

1.From the Assay Results (Figure 14) or All Events (Figure 16) screen, select record(s) by touching the desired record(s). Touch Transfer to open the Results or Events Transfer screen (Figure 18 shows results transfer as an example).

IQC and LQC results or failures can be transferred or printed in the same manner from their respective screens.

2.On the Results or Events Transfer screen, touch Print, USB or Network as desired. On the Events Transfer screen, touch USB or Network as desired. Only options for which the connections are available will be functional.

3.Confirm the transfer by touching Confirm. Touch Cancel to abort transfer. (Figure 19)

When the transfer is complete, touch OK and the Results screen (or Events screen) will be displayed. If transfer fails the warning icon and a failure message are displayed, touch OK to return to Results or Events screen (the screen depends on the previous context, e.g., if transferring LQC results, then LQC Results screen will be shown).

Once results have been transferred, File Converter 1_1.xls on the supplied USB flash drive may be used to convert the data files to an Excel format.

Figure 18. Results Transfer

Figure 19. Result Transfer Confirmation

Open the file “File Converter 1_1.xls”, enable Macros, and follow the directions on the screen. You will be prompted to save the Excel-formatted file.

NOTE: File Converter 1_1.xls must be used for this purpose. Data will not be formatted properly if an earlier version of File Converter.xls is used.

17

Image 22
Contents Operator Manual Made in Canada for Table of Contents Maintenance Iii Read all contents of this manual prior to use Consult Accompanying Documents Consult Operator ManualTo reduce the risk associated with hazardous voltage Summary of Warnings and PrecautionsTo reduce the risk of instrument or accessory damage To reduce the risk associated with incorrect resultsOverview IntroductionIntended Use Optional Accessories Available from 3M Unpacking and Setup3M Rapid Detection Reader Components Barcode Scanner 3M Rapid Detection Reader SetupTo set-up 3M Rapid Detection Reader § 3M Rapid Detection Control Module CM3M Rapid Detection Reader 3M Rapid Detection Control Module 3M Rapid Detection Test Module CM Port3M Rapid Detection Reader Modules Reader Configuration Utility RCU supplied with the CM. ForTurning on the 3M Rapid Detection Reader Display Icons Port Identification NomenclaturePerforming an Assay Using Lot CardsStarting an Assay Operator Entry Running Quality Control AcceptLQC Icons Assay Status WarningsTo run IQC manually Running IQCTo run IQC automatically Viewing Saved Results Working with Results and LogsViewing Saved Logs Searching Results Searching LogsManually Transferring / Printing Results or Logs Results TransferManaging Settings LQCSettings View Tabs AssayTimer Login PrinterFile Login RequirementsNetwork Settings Privilege RequirementsSettings View About Tab AboutSettings Edit Tabs Edit SettingsTimers Language Manual / Auto Transfer ModeDate/Time MiscManaging Operators Restoring DefaultsSound Adding/Modifying an Operator To Add an OperatorSearching for an Individual Operator Touch Search or CancelTo Modify an Operator Login Editing Settings Using the Reader Configuration Utility RCUInstallation Getting StartedSetting Up a USB Flash Drive Emergency Password Reset MechanismChoosing a Configuration File Access Method Setting Up Web Services AccessAccessing Configuration Files To open all files for a groupTo open an Individual Configuration File Setting Up Communications Between RCU and ReadersFiles Networked Environment Identifying Readers through the RCUNon-networked Environment Touch Export to USB Flash DriveImporting Files Exporting FilesDeleting Results Delete Results Prior to DateExterior Cleaning MaintenancePossible device damage Upgrading Device Software Disposal of EquipmentRestoring the Supervisor Operator ID / Password TroubleshootingGeneral No Available Ports. IQC Failed To Start Power On Self Test Message / IQC MessagesTest Run Messages FailError Insertion Error Remove & Reinsert Followed by Timed OutLot Card Messages Settings Messages LQC Related Messages Printer Errors Communications Errors USB Device ErrorsUnrecognized USB Drive Attached, Remove Device Specifications Operating Humidity Operating temperatureStorage temperature Operating/StorageIndex 3M Rapid Detection ReaderRestoring Supervisor Operator ID Password Languages Upgrading Software See RCU USB