3M #3960, #3968 manual Deleting Results, Delete Results Prior to Date

Page 40

3M™ Rapid Detection Reader

Editing Settings Using the Reader Configuration Utility (RCU)

Deleting Results

The 3M Rapid Detection Reader can store up to 900 results (300 patient, 300 LQC and 300 IQC) and 500 events (log file entries). When the CM database is full, the oldest result is over-written.

In Manual Transfer mode only (i.e., when 3M Rapid Detection Reader is not connected to a network), the warning “Database Almost Full, Backup Data” will be displayed when the patient or QC result databases have less than 20 spaces left. No warning is displayed when the event log is full.

NOTE: Prior to deleting results, transfer and store all data in accordance with applicable local, state and/or federal guidelines or accreditation requirements and each facility’s protocol.

In Auto Transfer mode, only new results that have not been previously transferred are transferred during an upload. Data is not deleted from CM upon transfer.

1.From the Settings menu (Figure 20), touch Delete Results. The Delete Results screen will open (Figure 47).

2.Touch Delete All Results or Delete Results Prior to Date. Delete All Results:

Touch Confirm.

Delete Results Prior to Date:

Enter the desired date (Figure 48), touch Confirm.

Once Results Deleted message is displayed, touch OK to return to the Settings menu.

Figure 47. Settings: Delete Results

Figure 48. Settings: Delete Results Prior to Date

35

Image 40
Contents Operator Manual Made in Canada for Table of Contents Maintenance Iii Read all contents of this manual prior to use Consult Accompanying Documents Consult Operator ManualSummary of Warnings and Precautions To reduce the risk of instrument or accessory damageTo reduce the risk associated with hazardous voltage To reduce the risk associated with incorrect resultsOverview IntroductionIntended Use Optional Accessories Available from 3M Unpacking and Setup3M Rapid Detection Reader Components 3M Rapid Detection Reader Setup To set-up 3M Rapid Detection ReaderBarcode Scanner § 3M Rapid Detection Control Module CM3M Rapid Detection Reader 3M Rapid Detection Control Module 3M Rapid Detection Test Module CM Port3M Rapid Detection Reader Modules Reader Configuration Utility RCU supplied with the CM. ForTurning on the 3M Rapid Detection Reader Display Icons Port Identification NomenclaturePerforming an Assay Using Lot CardsStarting an Assay Operator Entry Running Quality Control AcceptLQC Icons Assay Status WarningsTo run IQC manually Running IQCTo run IQC automatically Viewing Saved Results Working with Results and LogsViewing Saved Logs Searching Results Searching LogsManually Transferring / Printing Results or Logs Results TransferManaging Settings LQCSettings View Tabs AssayTimer Login PrinterLogin Requirements Network SettingsFile Privilege RequirementsSettings View About Tab AboutSettings Edit Tabs Edit SettingsTimers Manual / Auto Transfer Mode Date/TimeLanguage MiscManaging Operators Restoring DefaultsSound Adding/Modifying an Operator To Add an OperatorSearching for an Individual Operator Touch Search or CancelTo Modify an Operator Editing Settings Using the Reader Configuration Utility RCU InstallationLogin Getting StartedEmergency Password Reset Mechanism Choosing a Configuration File Access MethodSetting Up a USB Flash Drive Setting Up Web Services AccessAccessing Configuration Files To open all files for a groupTo open an Individual Configuration File Setting Up Communications Between RCU and ReadersFiles Identifying Readers through the RCU Non-networked EnvironmentNetworked Environment Touch Export to USB Flash DriveImporting Files Exporting FilesDeleting Results Delete Results Prior to DateExterior Cleaning MaintenancePossible device damage Upgrading Device Software Disposal of EquipmentRestoring the Supervisor Operator ID / Password TroubleshootingGeneral Power On Self Test Message / IQC Messages Test Run MessagesNo Available Ports. IQC Failed To Start FailError Insertion Error Remove & Reinsert Followed by Timed OutLot Card Messages Settings Messages LQC Related Messages Printer Errors Communications Errors USB Device ErrorsUnrecognized USB Drive Attached, Remove Device Specifications Operating temperature Storage temperatureOperating Humidity Operating/StorageIndex 3M Rapid Detection ReaderRestoring Supervisor Operator ID Password Languages Upgrading Software See RCU USB