3M #3960 Identifying Readers through the RCU, Non-networked Environment, Networked Environment

Page 38

3M™ Rapid Detection Reader

Editing Settings Using the Reader Configuration Utility (RCU)

Identifying Readers through the RCU

The RCU uses Reader information files to identify the Readers installed within an institution. Once a Reader has been identified, the RCU may be used to change Reader specific settings.

Non-networked Environment

A USB flash drive is used to transfer Reader settings between the Reader and the RCU. It is recommended that the USB flash drive provided with the 3M Rapid Detection Reader is used as it is pre-configured with the appropriate folders.

1.Insert a USB flash drive in an available USB port on the CM. Ensure that the USB drive contains four folders named: lqc, patient, iqc, log.

2.Navigate to the Reader Settings screen and touch Export Configuration.

3.Touch Export to USB Flash Drive.

4.All settings files and result logs are exported.

5.Steps 1 to 3 can be repeated for additional Readers, if desired.

6.Insert the USB flash drive into the RCU computer’s USB drive.

7.From the RCU, click on File > Open, select USB Device (navigate to the correct location of the USB Device if this function does not work automatically). Select the Files of Type

– Readers. Select the desired Reader Serial number and then click Open.

8.Changes to settings for the selected Reader may now be defined on the Reader Settings menu.

For more information see Importing Files or Exporting Files below.

Networked Environment

Each Reader must be placed into network mode and must be assigned a device information directory prior to successful bi-directional communications.

1.Using the Reader Touch screen, navigate to the Reader Settings: Edit Settings: File screen (Figure 29).

2.Enter the Device and Group Directory Settings for the Reader as defined in the RCU.

3.Navigate to the Reader Settings: Edit Settings: Misc screen (Figure 32) and enable Auto Transfer mode.

4.From the RCU, click on File > Open, and then select Auto Transfer Mode. Select the Files of Type – Readers. Browse to the device directory for the Reader of interest, select the desired Reader serial number and then click Open.

Settings for the selected Reader may now be defined on the Reader Settings menu.

For detailed information on all the RCU functions, refer to the Help Menu in the RCU program.

33

Image 38
Contents Operator Manual Made in Canada for Table of Contents Maintenance Iii Read all contents of this manual prior to use Consult Accompanying Documents Consult Operator ManualTo reduce the risk associated with hazardous voltage Summary of Warnings and PrecautionsTo reduce the risk of instrument or accessory damage To reduce the risk associated with incorrect resultsIntended Use IntroductionOverview 3M Rapid Detection Reader Components Unpacking and SetupOptional Accessories Available from 3M Barcode Scanner 3M Rapid Detection Reader SetupTo set-up 3M Rapid Detection Reader § 3M Rapid Detection Control Module CM3M Rapid Detection Reader 3M Rapid Detection Control Module 3M Rapid Detection Test Module CM PortTurning on the 3M Rapid Detection Reader Reader Configuration Utility RCU supplied with the CM. For3M Rapid Detection Reader Modules Display Icons Port Identification NomenclatureStarting an Assay Using Lot CardsPerforming an Assay Operator Entry Running Quality Control AcceptLQC Icons Assay Status WarningsTo run IQC automatically Running IQCTo run IQC manually Viewing Saved Logs Working with Results and LogsViewing Saved Results Searching Results Searching LogsManually Transferring / Printing Results or Logs Results TransferManaging Settings LQCSettings View Tabs AssayTimer Login PrinterFile Login RequirementsNetwork Settings Privilege RequirementsSettings View About Tab AboutTimers Edit SettingsSettings Edit Tabs Language Manual / Auto Transfer ModeDate/Time MiscSound Restoring DefaultsManaging Operators Adding/Modifying an Operator To Add an OperatorTo Modify an Operator Touch Search or CancelSearching for an Individual Operator Login Editing Settings Using the Reader Configuration Utility RCUInstallation Getting StartedSetting Up a USB Flash Drive Emergency Password Reset MechanismChoosing a Configuration File Access Method Setting Up Web Services AccessAccessing Configuration Files To open all files for a groupFiles Setting Up Communications Between RCU and ReadersTo open an Individual Configuration File Networked Environment Identifying Readers through the RCUNon-networked Environment Touch Export to USB Flash DriveImporting Files Exporting FilesDeleting Results Delete Results Prior to DatePossible device damage MaintenanceExterior Cleaning Upgrading Device Software Disposal of EquipmentGeneral TroubleshootingRestoring the Supervisor Operator ID / Password No Available Ports. IQC Failed To Start Power On Self Test Message / IQC MessagesTest Run Messages FailError Insertion Error Remove & Reinsert Followed by Timed OutLot Card Messages Settings Messages LQC Related Messages Printer Errors Communications Errors USB Device ErrorsUnrecognized USB Drive Attached, Remove Device Specifications Operating Humidity Operating temperatureStorage temperature Operating/StorageIndex 3M Rapid Detection ReaderRestoring Supervisor Operator ID Password Languages Upgrading Software See RCU USB