3M™ Rapid Detection ReaderTroubleshooting
Settings Messages
Message | Cause | Action |
|
|
|
User Expired, Try Again
User certification has expired or the details were entered incorrectly.
LQC Related Messages
Message |
| Cause |
|
| Action |
|
LQC Lockout. Ejecting. |
| LQC is overdue for this assay |
|
| LQC must be run and passed, |
|
|
| and no overrides are available. |
|
| or a Full Access User can reset |
|
|
|
|
|
| the LQC timer and/or overrides |
|
|
|
|
|
| for this assay in Reader |
|
|
|
|
|
| Settings on the RCU. |
|
|
|
|
|
|
|
|
Invalid Control. Ejecting. |
| Control ID does not exist for this |
|
| Repeat LQC with correct |
|
|
| assay or has been entered |
|
| Control ID or contact |
|
|
| incorrectly. |
|
| Supervisor. |
|
|
|
|
|
|
|
|
LQC Overdue, Assays Locked |
| LQC has expired and all |
|
| Run the required LQC. |
|
|
| available overrides have been |
|
|
|
|
|
| used. Tests cannot be run until |
|
|
|
|
|
| the required LQC has been run |
|
|
|
|
|
| and passed. |
|
|
|
|
|
|
|
|
|
|
|
|
| LQC will be due shortly. |
|
| Run the required LQC. |
|
|
| Appears when the LQC timer for |
|
|
|
|
|
| an assay has less than the user |
|
|
|
|
|
| defined % of time remaining until |
|
|
|
|
|
| lockout. |
|
|
|
|
|
|
|
|
|
|
|
|
| LQC has expired for one or more |
|
| Run the required LQC. |
|
|
| assays. If icon is displayed, the |
|
| A limited number of overrides |
|
|
| Assay Status Warnings screen |
|
|
| |
|
|
|
| can be configured via the RCU. |
| |
|
| (Figure 11) will display upon |
|
|
| |
|
|
|
| If overrides are available, touch |
| |
|
| touching the Run Assay button. |
|
|
| |
|
|
|
| Continue to override lock out |
| |
|
|
|
|
|
| |
|
|
|
|
| (i.e., perform a patient test after |
|
|
|
|
|
| LQC is due). The decision to |
|
|
|
|
|
| use an Override will be |
|
|
|
|
|
| confirmed (Figure 12). Once all |
|
|
|
|
|
| overrides have been used, LQC |
|
|
|
|
|
| must be performed prior to any |
|
|
|
|
|
| further patient testing. If an |
|
|
|
|
|
| assay with no overrides is |
|
|
|
|
|
| inserted into the TM, an LQC |
|
|
|
|
|
| Overdue, Assay Locked |
|
|
|
|
|
| warning will display and the |
|
|
|
|
|
| assay will be aborted. |
|
|
|
|
|
|
|
|
43