3M #3968, #3960 manual Setting Up Communications Between RCU and Readers, Files

Page 37

3M™ Rapid Detection Reader

Editing Settings Using the Reader Configuration Utility (RCU)

To open an Individual Configuration File

If only one configuration file requires modification, for example, when updating Operator Lists, there is no need to open all the Group files, just the file to be changed (in this example, userlist.opl). To do this, click on File > Open, and then select Defaults..., Network... or USB Device....

A “File Open” screen will display allowing navigation to the drive containing the files of interest (Figure 41).

To select different types of files to load, select the Files of Type dropdown list. As each file type is selected, a list of those files present in the currently selected directory will be made available. Select the desired file name and then click Open.

Figure 41. Open Default Configuration

Setting Up Communications Between RCU and Readers

In the RCU: Open the default Group Settings file (Figure 42).

Files

This page sets the file locations where the Reader(s) looks for specific configuration information. The Reader supports the HTTP, HTTPS, FTP and FTPS file transfer protocols. Directories are specified in URL notation. Please consult the local IT administrator for guidance on the use of these settings in the local network environment. If Readers will use the USB flash drive for configuration information, all file location fields should be set to: file:/// as in the default settings (except time server, which should be blank).

Examples of supported configurations are listed below.

USB Flash drive: "file:/// "

Web Server: "http://medsvr/3M Rapid Detection Reader/ "

Network drive: "ftp://fileserver/3M Rapid Detection Reader_share/"

The Group Directory field identifies where the Reader(s) looks for all group specific information.

The Assay List field identifies where the Reader(s) looks for its list of supported assays.

The LQC Controls List field identifies where the Reader(s) looks for its list of LQC controls.

The Operator Accounts List field identifies where the Reader(s) looks for its operator list.

The Results Directory field identifies where the Reader(s) places exported result files. In a networked environment one or more groups can share the same destination directory.

Figure 42. Reader Configuration: Group Settings

The Device Directory field identifies where the Reader(s) should read and write device specific information – this is typically set to the same directory as the Group Directory.

The Time Server Location allows the Reader group to access an NTP (Network Time Protocol) server to keep the Reader's internal clocks set to the correct time. This field should be left blank in non-networked environments or networked environments that lack access to an NTP server.

32

Image 37
Contents Operator Manual Made in Canada for Table of Contents Maintenance Iii Consult Accompanying Documents Consult Operator Manual Read all contents of this manual prior to useTo reduce the risk of instrument or accessory damage Summary of Warnings and PrecautionsTo reduce the risk associated with hazardous voltage To reduce the risk associated with incorrect resultsOverview IntroductionIntended Use Optional Accessories Available from 3M Unpacking and Setup3M Rapid Detection Reader Components To set-up 3M Rapid Detection Reader 3M Rapid Detection Reader SetupBarcode Scanner § 3M Rapid Detection Control Module CM3M Rapid Detection Reader 3M Rapid Detection Test Module CM Port 3M Rapid Detection Control Module3M Rapid Detection Reader Modules Reader Configuration Utility RCU supplied with the CM. ForTurning on the 3M Rapid Detection Reader Port Identification Nomenclature Display IconsPerforming an Assay Using Lot CardsStarting an Assay Operator Entry Accept Running Quality ControlAssay Status Warnings LQC IconsTo run IQC manually Running IQCTo run IQC automatically Viewing Saved Results Working with Results and LogsViewing Saved Logs Searching Logs Searching ResultsResults Transfer Manually Transferring / Printing Results or LogsLQC Managing SettingsAssay Settings View TabsTimer Printer LoginNetwork Settings Login RequirementsFile Privilege RequirementsAbout Settings View About TabSettings Edit Tabs Edit SettingsTimers Date/Time Manual / Auto Transfer ModeLanguage MiscManaging Operators Restoring DefaultsSound To Add an Operator Adding/Modifying an OperatorSearching for an Individual Operator Touch Search or CancelTo Modify an Operator Installation Editing Settings Using the Reader Configuration Utility RCULogin Getting StartedChoosing a Configuration File Access Method Emergency Password Reset MechanismSetting Up a USB Flash Drive Setting Up Web Services AccessTo open all files for a group Accessing Configuration FilesTo open an Individual Configuration File Setting Up Communications Between RCU and ReadersFiles Non-networked Environment Identifying Readers through the RCUNetworked Environment Touch Export to USB Flash DriveExporting Files Importing FilesDelete Results Prior to Date Deleting ResultsExterior Cleaning MaintenancePossible device damage Disposal of Equipment Upgrading Device SoftwareRestoring the Supervisor Operator ID / Password TroubleshootingGeneral Test Run Messages Power On Self Test Message / IQC MessagesNo Available Ports. IQC Failed To Start FailInsertion Error Remove & Reinsert Followed by Timed Out ErrorLot Card Messages Settings Messages LQC Related Messages Printer Errors USB Device Errors Communications ErrorsUnrecognized USB Drive Attached, Remove Device Specifications Storage temperature Operating temperatureOperating Humidity Operating/Storage3M Rapid Detection Reader IndexRestoring Supervisor Operator ID Password Languages Upgrading Software See RCU USB