3M™ Rapid Detection ReaderTroubleshooting
Test Run Messages
Message / Symptom  | 
  | Cause  | 
  | 
  | Action  | 
  | 
Barcode Error Aborting  | 
  | The 3M Rapid Detection  | 
  | 
  | Repeat assay using a fresh set of  | 
  | 
Assay  | 
  | Reader was unable to read the  | 
  | 
  | components.  | 
  | 
  | 
  | barcode on the Test Cartridge.  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
No Ports Available, Assays  | 
  | All ports are in use by other  | 
  | 
  | Touch OK to return to the main  | 
  | 
in Progress  | 
  | assays.  | 
  | 
  | menu and wait until an assay is  | 
  | 
  | 
  | 
  | 
  | 
  | complete before attempting to run  | 
  | 
  | 
  | 
  | 
  | 
  | another test.  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
No Ports Available, Remove  | 
  | Cartridges occupying all ports  | 
  | 
  | Remove Test Cartridge(s) for  | 
  | 
Cartridge(s) Try Again  | 
  | 
  | 
  | 
  | completed assay(s).  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
Lot Card Messages | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
Message / Symptom  | 
  | Cause  | 
  | 
  | Action  | 
  | 
No response to insertion of a  | 
  | Lot Card Assembly not  | 
  | 
  | Contact Technical Support to  | 
  | 
Lot Card.  | 
  | functioning.  | 
  | 
  | report the problem.  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
Lot Card Error, Reinsert | 
  | The 3M Rapid Detection  | 
  | 
  | Touch OK, then reinsert the Lot  | 
  | 
  | 
  | Reader was unable to read the  | 
  | 
  | Card.  | 
  | 
  | 
  | Lot Card correctly.  | 
  | 
  | If the problem persists use a Lot  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | |
  | 
  | 
  | 
  | 
  | Card from another box of Tests  | 
  | 
  | 
  | The Lot Card may have been  | 
  | 
  | with the same Lot number.  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | |
  | 
  | removed too quickly, or not  | 
  | 
  | If another Lot Card is not available  | 
  | 
  | 
  | fully inserted into the Lot Card  | 
  | 
  | or alternate lot card fails, contact  | 
  | 
  | 
  | slot.  | 
  | 
  | Technical Support.  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
Lot # XXX Not Found, Insert  | 
  | The 3M Rapid Detection  | 
  | 
  | Insert the Lot Card for the  | 
  | 
Lot Card or Cancel | 
  | Reader does not contain the  | 
  | 
  | specified lot or cancel assay.  | 
  | 
  | 
  | data required to run Test  | 
  | 
  | NOTE: The correct Lot Card must  | 
  | 
  | 
  | Cartridges with that Lot #  | 
  | 
  | 
  | |
  | 
  | 
  | 
  | be inserted within approximately  | 
  | |
  | 
  | 
  | 
  | 
  | 
  | |
  | 
  | The Lot Card for this Test Kit  | 
  | 
  | 30 seconds or the assay will be  | 
  | 
  | 
  | Lot may not have been  | 
  | 
  | aborted.  | 
  | 
  | 
  | inserted before running the  | 
  | 
  | Repeat assay using a fresh set of  | 
  | 
  | 
  | assay.  | 
  | 
  | 
  | |
  | 
  | 
  | 
  | components.  | 
  | |
  | 
  | 
  | 
  | 
  | 
  | |
  | 
  | 
  | 
  | 
  | 
  | 
  | 
Lot Card Expired | 
  | The Lot Card inserted has  | 
  | 
  | Check the date/time settings  | 
  | 
  | 
  | expired. Expiry falls on the last  | 
  | 
  | within the 3M Rapid Detection  | 
  | 
  | 
  | day of the month.  | 
  | 
  | Reader.  | 
  | 
  | 
  | 
  | 
  | 
  | Use components that have not  | 
  | 
  | 
  | 
  | 
  | 
  | expired.  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
41