Sunrise Medical 5650K, 5650P, 5650H, 3655D Specifications, Specification s, Electrical, Operating

Page 12

12

S p e c i f i c a t i o n s

SPECIFICATIONS

Size 3655 Series

7.5" W x 4.0" H x 7.2" D (19.1 cm x 10.2 cm x 18.3 cm)

4650 Series

7" W x 3.8" H x 13" D (18 cm x 10 cm x 33 cm)

5650 Series

10.1" W x 10.5" H x 6.5" D (26 cm x 27 cm x 17 cm)

Weight 3655 Series

4.8 lbs. (2.18 kg)

4650 Series

5.3 lbs. (2.40 kg)

5650 Series

7.1 lbs. (3.2 kg)

Electrical

 

 

Requirements 3655D, 3655T, 4650D & 46502

2.5 amps, 115 VAC (±10%) 60 Hz

3655I & 4650I

1.4 amps, 230 VAC (±10%) 50 Hz

3655K & 3655P

1.4 amps, 230 VAC (±10%) 60 Hz

5650D, 5650H & 5650C

1.3 amps, 115VAC 60Hz

5650I

0.8 amps, 230VAC 50Hz

5650K & 5650P

0.8 amps, 230VAC 50/60Hz

Sound Level 3655 Series

58 dBA

4650 Series

53 dBA

5650 Series

51 dBA

Power Consumption 3655D, 3655T, 4650D & 46502

140 watts maximum

3655I, 3655K, 3655P & 4650I

160 watts maximum

5650D, 5650H & 5650C

90 watts maximum

5650I, 5650K & 5650P

100 watts maximum

Maximum Compressor Pressure 3655 & 4650 series

35 psig or greater (241 kPa)*

 

5650 series

30 psig or greater (206 kPa)*

Compressor Free Air Flow 3655 & 4650

8 lpm or greater*

Operating

5650

9 lpm or greater*

 

 

Temperature Range all Models except 5650C

+40° to +104°F (+5 to +40°C)

5650C

+40° to +101°F (+5 to +38.5°C)

Operating Humidity all Models

up to 95% non-condensing

Storage/Transport Temperature Range all Models

-40° to +158°F (-40° to +70°C)

Storage/Transport Humidity all Models

up to 95% non-condensing

Supplied Nebulizer Capacity all models

6 ml (cc)

Supplied Nebulizer Nebulization Rate all models

0.15 ml/min or greater*

Supplied Nebulizer MMAD all models

5 microns or less**

*Conditions may vary based on altitude above sea level, barometric pressure, and temperature. **Value determined with minimum flow rate through the nebulizer (20°C, sea level).

There are no hazardous materials used in the construction of this unit, therefore, there are no known risks associated with disposal of the equipment.

This product (models 3655D & 3655T) complies with the following electromagnetic compatibility standard: IEC 60601-1-2:2nd Edition

3655D & 3655T Equipment Classifications:

 

Protection against electric shock:

Class II

Degree of protection against electric shock:

Type BF

Equipment not suitable for use in the presence of flammable anaesthetic mixture with

air or with oxygen or nitrous oxide

 

Mode of operation:

Continuous operation

A-655

Image 12
Contents DeVilbiss Compressor/Nebulizer Instruction Guide English EspañolFrançais B l e N t e n t sP o r t a n t F e g u a r d s To reduce the risk of electrocutionModel 5650H & 4650D Grounding Instructions Introduction P o r t a n t F e g u a r d s T r o d u c t i o nStatement of Intended Use P o r t a n t R t s U r M p r e s s o r Figure a On backHOW to Operate Your Compressor P o r t a n t R t s U r B u l i z e rModels 4650 & 5650 Steps 1-3 and Steps E r a t e U r M p r e s s o r Models 3655 Steps 1-3 and StepsNebulizer Cleaning E a n i n gCompressor Cleaning Disinfect DailyMaintenance Filter ChangeL t e r A n g e I n t e n a n c e Models 4650Specifications Specification sElectrical OperatingFIVE-YEAR Limited Warranty on Compressor Parts Manufacturers NoteR r a n t y LEA Todas LAS Instrucciones Antes DE Usarlo E c a u c i o n e s P o r t a n t e sPara reducir el riesgo de electrocutación Peligro Introducción Enunciado de uso idóneoR t e s P o r t a n t e s Figuras a y BPartes Importantes DE SU Nebulizador Desechable Partes Importantes DE SU Nebulizador ReutilizableCómo Operar SU Compresor A r t e s P o r t a n t e sM o E r a r M p r e s o r Modelos 3655 Pasos 1-3 y pasosTodos Modelos Pasos Limpieza DEL Nebulizador M p i e z aPara evitar un posible riesgo de infección debido a Un medicamento contaminado, se recomiendaLimpieza DEL Compresor DiariamenteNo lo sumerja en agua, ya que dañaría el compresor Cambio DE Filtro MantenimientoM b i o L t r o N t e n i m i e n t o Modelos 4650 yEspecificaciones P e c i f i c a c i o n e sNota DEL Fabricante R a n t í aLire Toutes LES Instructions Avant L’UTILISATION S e s R d e P o r t a n t e sPour réduire les risques d’électrocution S e s R d e P o r t a n t e s Garder CES Instructions Déclaration d’utilisation envisagéeI è c e s P o r t a n t e s Figures a et BPièces Importantes DE Votre Nébuliseur Jetable Pièces Importantes DE Votre Nébuliseur RéutilisableÈ c e s P o r t a n t e s Modèles 4650 et 5650 étapes 1-3 et étapesM m e n t I r e N c t i o n n e r M p r e s s e u r Nettoyage DU Nébuliseur T t o y a g eNettoyage DU Compresseur Désinfecter chaque joursChangement DU Filtre EntretienA n g e m e n t L t r e T r e t i e n Modèles 4650 etSpécifications É c i f i c a t i o n sFlux d’air libre du compresseur- Séries 3655 et Amplitude de températuresR a n t i e Classes d’équipement de 3655D et 3655TSunrise Medical Sunrise Medical Canada, Inc