Sigma S2Ca, S2Ba Settings for Òauxiliary frequencyÓ function AUX menu, Setting code Code menu

Page 51

Dulcodes UV-Desinfektionsanlage

Settings

 

 

7.5.2Settings for Òauxiliary frequencyÓ function (AUX menu)

￿￿￿

Continuous display

P

Set ￿￿￿

 

Aux￿￿

 

P

Set ￿

Freq.

P

The programmable function “auxiliary frequency” allows switching to a different stroke frequency, which can be set in the AUX menu. It can be activated via the “external control” terminal. When the auxiliary frequency is activated, “aux” appears in the LCD display.

This auxiliary frequency overrides the current stroke frequency set for the selected operating mode.

7.5.3Settings for the ÒflowÓ function (FLOW menu)

￿￿￿

Continuous display

P

P

Set ￿￿￿￿

Flow

Set ￿￿

P

Flow ￿￿￿

Set

P

The flow menu only appears when a dosing monitor is connected to the “dosing monitor” terminal. This dosing monitor registers each discharge stroke of the S2Ca at the discharge connector and transmits it back to the S2Ca. If this response transmission is serially omitted for a period set in the FLOW menu (due to failure or below-minimum dosing) the S2Ca stops.

7.6Setting code (CODE menu)

The code menu is used to select whether you want to prevent access to parts of the settings options.

￿￿￿￿ P ￿￿￿￿￿ P ￿￿￿￿

P

Continuous display

In the first menu option you can choose CODE 1 or CODE 2 (both use the same number).

Select CODE 1 to prevent access to the settings mode (in the overview “operating/ settings diagram”, fold-out page). In the next menu option, enter the number you wish to use as the code.

Select CODE 2 to prevent access to the settings options for directly alterable values in the continuous displays (in the overview “operating/settings diagram”, fold-out page). In the next menu option, enter the number you wish to use as the code.

Select NONE to remove a pre-set security lock.

7.7Cancel total stroke number or total litres (CLEAR window)

In the CLEAR window you can delete the stored total stroke number and simultaneously the total litres (= set to “0”). You may then press the P key briefly to exit this window.

The values displayed are counted incrementally from the point of commissioning the pump, or from the last delete action.

N

￿￿￿￿￿L

Continuous display

P

ProMinent®

Page 51

Image 51
Contents Operating Instructions Control elements overview Control elements and key functionsContinuous Display Operating-/Settings DiagramPrevious Activatedfunction AnalogImprint Imprint Dulcodes UV-DesinfektionsanlageDulcodes UV-Desinfektionsanlage Table of contents Table of contents Dulcodes UV-Desinfektionsanlage B a G m a B a s i c Ty p e S 2 B a Identcode ordering systemIdentcode ordering system Dulcodes UV-Desinfektionsanlage ProMinent Dulcodes UV-Desinfektionsanlage General notes Safety devicesDulcodes UV-Desinfektionsanlage Product description S2Ba/S2Ca Identification of pump typeDesign/Functional description Functional description, drive 75 % 30 % Stroke movementHM50Hz MeteringcapacitydiagramS2Ba Metering capacity diagramS2CaHM Functional description, delivery unit Integrated overflow valve with bleeder function TaskDesign and functional description refer to Fig. No Product description S2Ba/S2Ca Design and function description see Fig Diaphragm rupture sensorUse for intended purpose/Use unintended purpose Function3162-4 Pvdf Ptfe Technical data Sigma Capacity data Technical dataTechnical data S2Ca Temperature specificationsAccuracy Dimensions Sigma/ 2 mm Dimension sheet Sigma/ 2 with stroke adjustment motorModel Connector Electrical data, stroke sensor ÒSigmaÓ Stroke adjuster drive mechanismMotor data Stroke actuator drive mechanismRelay input Input voltage Mains frequency Power consumptionElectrical data, pacing relay ÒSigmaÓ Sound intensity levelStart-up/Maintenance MaintenanceStart-up Maintenance intervals Replacement of wearing partsGuideline on Valve Installation Features of the S2Ca metering pumps Function description, motorsFunction description, controller LED indicator Operating indicator green ÒFloat switchÓ functionOptional Òfault-indicating relayÓ or Òpower relayÓ ÒFault indicating and pacing relayÓ optionHierarchy of operating modes, functions and fault statuses Sockets, symbols and wiring diagramRelay type Power relay Wiring diagram Technical data, external contact Controller type 0 see identity codeController type 1 see identity code Outputs Block circuit diagram S2CaInputs Pump, InsideBy customer Cable Circuit examples universal signal cableNot in Profibus version Delivery range Retrofitting relaysControl elements Set SettingsCheck adjustable values Change to settings modeSelect operating mode Mode menu Settings for operating mode SET menu Settings for ÒmanualÓ operating modeSettings for ÒanalogueÓ operating mode Analg menu Straight line SetLower band Settings for ÒcontactÓ operating mode Contct menuLower/upper band Lower band Upper band Alkali pump Acid pumpSet ÒMemoryÓ extension function Settings for ÒbatchÓ operating mode Batch menuSettings for ÒcalibrationÓ function Calib menu Settings for programmable functions SET menuCancel total stroke number or total litres Clear window Settings for Òauxiliary frequencyÓ function AUX menuSettings for the ÒflowÓ function Flow menu Setting code Code menuManual operation OperatingRemote control TroubleshootingError Messages Decommissioning and disposal Spare parts and accessories 1011036 ProMinentFor pumps without Ex protection For pumps with Ex protection Valve assy DN 15/PVT* Order No 6105-1040100-03 Order No Retrofit kit double diaphragm S2Ba/S2Ca Wiring diagram for stroke length adjuster Stroke actuator wiring diagram