Arizona Jerome 631-X operation manual Instrument Operation, LCD Codes

Page 10

4. INSTRUMENT OPERATION

LCD Codes

 

 

 

 

 

LCD CODE

EXPLANATION

 

 

000

Ready to sample

 

 

 

 

 

 

.000

No hydrogen sulfide reading

 

 

 

 

 

 

.8.8.8

Sensor saturated-regeneration needed (refer to page 12)

 

 

 

 

 

 

.H.H.H

Sensor regeneration in progress (.H.H.H flashes)

 

 

.L.L.L

Re-zero needed (refer to page 13)

 

 

 

 

 

 

.P.P.P

Power cord required or low line power, <100 VAC (or 200 VAC)

 

 

 

(see pages 16 and 17, Changing the Fuse, if .P.P.P remains on after

 

 

 

the cord is connected.)

 

 

.LO BAT

Recharge batteries (refer to page 17)

 

 

 

 

 

 

.E.E.E

Same as LO BAT, automatically shuts off

 

 

 

 

 

 

.HL

Very high concentration has been detected. Refer to your safety

 

 

 

policy for additional direction to confirm the concentrations."

 

 

DURING SAMPLING

 

 

 

.-

0-25% sensor saturation

 

 

 

 

 

 

.--

25-50% sensor saturation

 

 

 

 

 

 

.---

50-75% sensor saturation

 

 

-.---

75-100% sensor saturation

 

 

 

 

 

 

DURING SAMPLING, USING SURVEY MODE

 

 

-

Survey sampling (minus sign flashes continuously)

 

WHEN ZERO IS

Adjust to 0 only after sensor regeneration. It is normal for the

DEPRESSED

display to read H after sampling has started.

0

Zero, ready to sample

 

 

H

High, turn Zero potentiometer counterclockwise

 

 

L

Low, turn Zero potentiometer clockwise

Page 10 of 50

AZI Customer Service 800-528-7411 or 602-470-1414

Image 10
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Optional Accessories FeaturesIntroduction Applications Principle of OperationGas Flow Schematics Zero Air FilterAZI Customer Service 800-528-7411 or LCD Codes Instrument OperationDisplay to read H after sampling has started Daily Operations Sensor Regeneration Do not Interrupt this CycleZero Adjust Do not turn the Zero Adjust potentiometer between samplesSample Mode Accuracy at MID-RANGESurvey Mode Operating on Internal Battery Power Operating on AC Power or GeneratorExternal battery power Obtaining Maximum Battery Life Charging BatteriesMaintenance Preventive Maintenance CalendarPart Part Number Flow SystemInch Fritware Filter Internal Filters Replacing the Battery Pack Setting the Input VoltageChanging the Fuse Calibration Verification of Calibration and Quality ControlPower Problems TroubleshootingRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Optional Communications Capability Instrument I/O InterfaceAZI Customer Service 800-528-7411 or Accessories & Maintenance Parts Functional Test ModuleJerome Data Logger Jerome Communication Software KitSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Appendix a 631-X Functional Test Module Simple OperationInstrument Control with Real-Time Graphics Appendix B Jerome Communications SoftwareMain System Requirements JCI Kit ContentsData Logger Option Switch # Position Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Function CommentAuto-Zero Appendix D Jerome 631-X Option BoardInstrument Zeroing Timed Regeneration MA Analog Output Auto-SampleSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Fresh Air Solenoid Required Parts Suggested Part Similar AZI P/NDC Power Operation DC Power Adaptor Kit, AZI P/N Y031AZI Customer Service 800-528-7411 or Molecular weight of H 2S Appendix E Calculation of Dynacalibrator H2S ConcentrationWarranty Trademark and Copyright Protection Arizona Instrument LLC Part Number January