Arizona Jerome 631-X Fresh Air Solenoid, Required Parts Suggested Part Similar AZI P/N

Page 45

Fresh Air Solenoid

An external three-way solenoid can be used to provide fresh air or conditioned air during sensor regeneration. This may be necessary if the sample stream lacks molecular oxygen. A low current six volt DC solenoid, connected between pins 19 and 11 of the 25 pin rear panel connector, will be energized during the regeneration cycle if the option board SW100 switch 6 is placed in the OFF position.

If needed, the circuit may be built from the following components and configured as shown in the following diagram. It will only function if the option board is installed in the 631-X instrument.

Required Parts:

Suggested Part

Similar AZI P/N

1 solenoid, 6volt 3way

Angar P/N 407569

1300 1004

1/8" to 1/16" tubing adaptor

Any

1300 0025

1/2" clamp, adhesive mount

Any

6000 0013

1/8" tube to instrument adaptor

Any

1400 3010

3" 1/8" tubing

Tygon Formula R3603

2500 3001

A/R 1/16" tubing

Tygon Formula R3603

2500 3002

1 25 pin male DB-25 connector

 

 

Solder-cup style

AMP 747912-2

None *

1 connector hood

AMP 749626-2

None *

*These are types not stocked by AZI, but should be available overnight from many AMP stocking distributors such as Digi-Key Corporation. There are multiple suitable alternatives such as Radio Shack’s 276-1547 and 276-1549.

 

 

 

1/8" TUBE

1/16" TUBE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1/16" TUBE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TO INSTRUMENT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TO SAMPLE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SOURCE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1/16" TUBE

TO FRESH AIR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SOURCE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PIN 1

 

 

 

 

 

 

 

 

 

PIN 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

REAR OF CONNECTOR,

 

 

 

 

REAR OF CONNECTOR,

 

 

CONNECT TO PINS 11 AND 19

 

CONNECT TO PINS 19 AND 21

 

MOST INSTRUMENTS

 

 

631-X SHIPPED IN 1994 ONLY

AZI Customer Service 800-528-7411 or 602-470-1414

Page 45 of 50

Image 45
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Features Optional AccessoriesIntroduction Principle of Operation ApplicationsZero Air Filter Gas Flow SchematicsAZI Customer Service 800-528-7411 or Instrument Operation LCD CodesDisplay to read H after sampling has started Daily Operations Do not Interrupt this Cycle Sensor RegenerationDo not turn the Zero Adjust potentiometer between samples Zero AdjustAccuracy at MID-RANGE Sample ModeSurvey Mode Operating on AC Power or Generator Operating on Internal Battery PowerExternal battery power Charging Batteries Obtaining Maximum Battery LifePreventive Maintenance Calendar MaintenanceFlow System Part Part NumberInch Fritware Filter Internal Filters Setting the Input Voltage Replacing the Battery PackChanging the Fuse Verification of Calibration and Quality Control CalibrationTroubleshooting Power ProblemsRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Instrument I/O Interface Optional Communications CapabilityAZI Customer Service 800-528-7411 or Functional Test Module Accessories & Maintenance PartsJerome Communication Software Kit Jerome Data LoggerSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Simple Operation Appendix a 631-X Functional Test ModuleAppendix B Jerome Communications Software Instrument Control with Real-Time GraphicsMain JCI Kit Contents System RequirementsData Logger Option DIP Switch Settings Appendix C Internal DIP Switch SettingsSwitch # Position Function Comment Switch # Function CommentAppendix D Jerome 631-X Option Board Auto-ZeroInstrument Zeroing Timed Regeneration Auto-Sample MA Analog OutputSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Required Parts Suggested Part Similar AZI P/N Fresh Air SolenoidDC Power Adaptor Kit, AZI P/N Y031 DC Power OperationAZI Customer Service 800-528-7411 or Appendix E Calculation of Dynacalibrator H2S Concentration Molecular weight of H 2SWarranty Arizona Instrument LLC Part Number January Trademark and Copyright Protection