| 7. |
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
| Symptom | Possible Cause |
| Solution | ||||
|
| Power Problems |
|
|
|
|
| ||
|
| Unit does not turn ON. Unit | Discharged battery or | Recharge battery for a minimum of | |||||
|
| turns on when power cord is |
|
| 14 hours. Refer to page 17. | ||||
|
| plugged in. LCD displays 000 | Dead battery. |
|
|
|
| ||
|
| when instrument is operating on | Replace battery. Refer to page 21. | ||||||
|
| AC power. |
|
|
|
|
| ||
|
| Unit does not turn on when | Open fuse. | Replace fuse. Refer to page 22. | |||||
|
| connected to AC power cord. | Insufficient power. | Use a voltmeter to verify there is | |||||
|
|
|
|
| |||||
|
|
|
|
|
|
| power to the AC outlet. | ||
|
|
|
|
| Internal component failure. | Call AZI Customer Service for | |||
|
|
|
|
|
| information at | |||
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
| ||
|
| Regeneration & Zero Problems |
|
|
|
|
| ||
|
| LCD displays .8.8.8. | Sensor saturated. | Do not attempt to adjust zero pot. | |||||
|
|
|
|
|
|
| The sensor must be regenerated. | ||
|
|
|
|
|
|
| Refer to page 12 for information. | ||
|
| LCD displays .L.L.L when | Changes in temperature. | Readjust zero pot. See page 13 for | |||||
|
| taking first sample. |
|
| information. | ||||
|
| LCD displays H at finish of | Internal contamination may | Remove and replace fritware filter, | |||||
|
| sensor regeneration when zero is | redeposit hydrogen sulfide from |
| intake filter disk, scrubber filters | ||||
|
| pressed. |
| flow system onto gold film |
| and Tygon tubing. Refer to “Flow | |||
|
|
|
|
| sensor. |
| System” on Page 19. | ||
|
|
|
|
|
| Check tubing for kinks or crimps. | |||
|
|
|
|
|
| Repeat regeneration cycle. Refer to | |||
|
|
|
|
|
|
| page 12. | ||
|
| Zero adjust pot cannot be | Pot not turned sufficiently. |
| 1. Turn zero adjust up to 20 times | ||||
|
| adjusted to 0. |
|
| to reach the end. Pot will | ||||
|
|
|
|
|
|
| “click” softly. | ||
|
|
|
|
|
|
| 2. If no “0”, turn pot slowly in | ||
|
|
|
|
|
|
| opposite direction until display | ||
|
|
|
|
| Sensor may be ruptured or pot |
| reads “0”. | ||
|
|
|
|
|
| 3. If still unchanged, call AZI | |||
|
|
|
|
| may be broken. |
| |||
|
|
|
|
|
| Customer Service at (800) 528- | |||
|
|
|
|
|
|
| |||
|
|
|
|
|
|
| 7411 or | ||
|
|
|
|
|
|
|
|
|
|
Page 24 of 50 | AZI Customer Service |