Arizona Jerome 631-X operation manual AZI Customer Service 800-528-7411 or

Page 5

Check the voltage setting (110 or 220 VAC) on the back of the instrument. Ensure that it is set to the correct voltage. If the pointer is not aligned to the local voltage, turn the selector to point to the correct voltage.

Perform a sensor regeneration by following these steps:

Connect the line cord between the connector on the back of the 631-X and an AC power outlet.

Press the ON switch and then press the REGEN button.

The instrument will begin a 10 minute regeneration cycle, indicated by .H.H.H flashing on the display. Do not interrupt this cycle. For a complete description of this process, see page 12.

If any error message, such as .P.P.P, appears on the display, see the “Troubleshooting” section beginning on page 24.

When regeneration is complete, zero the sensor by pressing the ZERO button and turning the zero adjust screw, located under the handle, until the display reads 0.

The instrument is now ready to sample.

To ensure the input to the instrument contains no hydrogen sulfide or mercaptans, use a Zero Air Filter, AZI P/N Z2600 3905. The Zero Air Filter cleans the air sample and should produce sample readings of less than 0.003 mg/m3. Therefore, use the filter to:

Equilibrate the instrument to temperatures that are higher or lower than the instrument. Sample with filter installed until the reading is below 0.003 mg/m3.

Identify contamination within the unit.

Confirm the presence of hydrogen sulfide when readings are elevated. Install filter and verify that the readings go down with filter installed.

When the instrument measures hydrogen sulfide, the zero display will be replaced with H.

CAUTION

Do not adjust the ZERO after the instrument has measured hydrogen sulfide or before the next regeneration. (Occasionally the display may drop to .L.L.L (indicating low) between the initial zeroing and the first sample. It is OK to readjust the ZERO if the instrument has not measured hydrogen sulfide.)

The instrument is designed for ambient air monitoring. DO NOT allow the probe or the instrument’s intake to be exposed to any liquid.

The instrument is not explosion proof.

Press the SAMPLE button to start a 10 second sampling cycle.

Perform sensor regeneration after each day’s testing.

Perform another sensor regeneration and re-zero the instrument before each day’s use.

Perform sensor regeneration after 30 days of storage or inactivity.

Call AZI Customer Service, at (800) 528-7411 from the United States and Canada or (602) 470-1414 if you have any questions. If you prefer, you may send e-mail to support@azic.com

AZI Customer Service 800-528-7411 or 602-470-1414

Page 5 of 50

Image 5
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Introduction FeaturesOptional Accessories Principle of Operation ApplicationsZero Air Filter Gas Flow SchematicsAZI Customer Service 800-528-7411 or Display to read H after sampling has started Instrument OperationLCD Codes Daily Operations Do not Interrupt this Cycle Sensor RegenerationDo not turn the Zero Adjust potentiometer between samples Zero AdjustAccuracy at MID-RANGE Sample ModeSurvey Mode External battery power Operating on AC Power or GeneratorOperating on Internal Battery Power Charging Batteries Obtaining Maximum Battery LifePreventive Maintenance Calendar MaintenanceInch Fritware Filter Flow SystemPart Part Number Internal Filters Setting the Input Voltage Replacing the Battery PackChanging the Fuse Verification of Calibration and Quality Control CalibrationRegeneration & Zero Problems TroubleshootingPower Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Instrument I/O Interface Optional Communications CapabilityAZI Customer Service 800-528-7411 or Functional Test Module Accessories & Maintenance PartsJerome Communication Software Kit Jerome Data LoggerSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Simple Operation Appendix a 631-X Functional Test ModuleMain Appendix B Jerome Communications SoftwareInstrument Control with Real-Time Graphics Data Logger Option JCI Kit ContentsSystem Requirements DIP Switch Settings Appendix C Internal DIP Switch SettingsSwitch # Position Function Comment Switch # Function CommentInstrument Zeroing Appendix D Jerome 631-X Option BoardAuto-Zero Timed Regeneration SW101 Functions Auto-SampleMA Analog Output Connection and Setup AZI Customer Service 800-528-7411 or Required Parts Suggested Part Similar AZI P/N Fresh Air SolenoidDC Power Adaptor Kit, AZI P/N Y031 DC Power OperationAZI Customer Service 800-528-7411 or Appendix E Calculation of Dynacalibrator H2S Concentration Molecular weight of H 2SWarranty Arizona Instrument LLC Part Number January Trademark and Copyright Protection