11. APPENDIX A - 631-X FUNCTIONAL TEST MODULE
The Jerome Hydrogen Sulfide Functional Test Module provides a fast and easy method of verifying that Jerome
The Functional Test Module is beneficial:
•In applications where frequent verification of functionality is required, such as ISO 9000 documentation,
•To verify proper instrument operation when unexpected readings are obtained in normal sampling,
•As part of a weekly maintenance routine, and
•To determine if analyzer calibration is needed.
Simple Operation
The Functional Test Module includes a permeation tube containing hydrogen sulfide. When activated, the test module releases this H2S from the permeation tube at a specific, known concentration. The H2S flows over the gold film sensor of the Jerome analyzer, which then measures the amount of exposure to the gas. The flow rate and temperature of this release are factory set to provide a concentration of approximately 250 ppb (0.250 ppm) " 20%. The user then compares the reading on the Jerome analyzer with the known concentration from the module. If the H2S level shown on the analyzer’s display falls within the expected range (approximately 250 ppb), the instrument is functioning properly. If the level is not in the expected range, it should be returned to the factory for
See page 30 for a list of the components that make up the FTM and its accessory kit.
The FTM, with the exception of the permeation tube, carries a limited
Page 36 of 50 | AZI Customer Service |