Arizona operation manual Appendix D Jerome 631-X Option Board, Auto-Zero, Instrument Zeroing

Page 40

14. APPENDIX D - JEROME 631-X OPTION BOARD

Proper use of this board requires that the base instrument be fully functional and set correctly for the intended operation.

Auto-Zero

With the option board installed, the 631-X has a limited auto-zero function. This function cannot be disabled and is transparent to the user. The instrument can be manually zeroed as described in “Zero Adjust” on page 13. However, if the instrument is to be operated by personnel not familiar with the procedure or if it is operated unattended, the auto-zero function should satisfactorily zero the unit after each sensor regeneration.

Instrument Zeroing

The Jerome 631-X has essentially three ways to zero the sensor reading before samples are taken if the option board is installed.

The instrument automatically re-zeroes between samples so that each sample is a unique reading. To take a sample, simply press the SAMPLE button.

The manually adjusted zero, using the switch on the top of the 631-X is used to re-establish a baseline between the reference and sensor gold films only after a sensor regeneration. This zero is manually adjusted by pressing the ZERO button and turning the potentiometer on the top of the instrument until the display reads 0. Adjust only after sensor regeneration; it is normal for H to be displayed after sampling.

The 631-X option board provides an auto-zero feature following regeneration that is invisible to the user.

Page 40 of 50

AZI Customer Service 800-528-7411 or 602-470-1414

Image 40
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Optional Accessories FeaturesIntroduction Applications Principle of OperationGas Flow Schematics Zero Air FilterAZI Customer Service 800-528-7411 or LCD Codes Instrument OperationDisplay to read H after sampling has started Daily Operations Sensor Regeneration Do not Interrupt this CycleZero Adjust Do not turn the Zero Adjust potentiometer between samplesSample Mode Accuracy at MID-RANGESurvey Mode Operating on Internal Battery Power Operating on AC Power or GeneratorExternal battery power Obtaining Maximum Battery Life Charging BatteriesMaintenance Preventive Maintenance CalendarPart Part Number Flow SystemInch Fritware Filter Internal Filters Replacing the Battery Pack Setting the Input VoltageChanging the Fuse Calibration Verification of Calibration and Quality ControlPower Problems TroubleshootingRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Optional Communications Capability Instrument I/O InterfaceAZI Customer Service 800-528-7411 or Accessories & Maintenance Parts Functional Test ModuleJerome Data Logger Jerome Communication Software KitSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Appendix a 631-X Functional Test Module Simple OperationInstrument Control with Real-Time Graphics Appendix B Jerome Communications SoftwareMain System Requirements JCI Kit ContentsData Logger Option Appendix C Internal DIP Switch Settings DIP Switch SettingsSwitch # Position Function Comment Switch # Function CommentAuto-Zero Appendix D Jerome 631-X Option BoardInstrument Zeroing Timed Regeneration MA Analog Output Auto-SampleSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Fresh Air Solenoid Required Parts Suggested Part Similar AZI P/NDC Power Operation DC Power Adaptor Kit, AZI P/N Y031AZI Customer Service 800-528-7411 or Molecular weight of H 2S Appendix E Calculation of Dynacalibrator H2S ConcentrationWarranty Trademark and Copyright Protection Arizona Instrument LLC Part Number January