Arizona Jerome 631-X Appendix C Internal DIP Switch Settings, Switch # Position Function Comment

Page 39

13. APPENDIX C - INTERNAL DIP SWITCH SETTINGS

The DIP switch identified as SW2 on the 631-X control board is located at the top center of the board.

CAUTION:

Switch the instrument’s power to OFF before opening the instrument

and changing DIP switch settings.

DIP Switch Settings:

Switch #

Position

Function

Comment

1

ON

AUTORANGE

DISABLED (SEE DIP SWITCHES 5 & 6)

1

OFF

AUTORANGE

ENABLED

2

X

NOT USED

 

3

X

NOT USED

 

4

ON

ZERO

DISPLAY:

 

00-99

 

 

 

 

 

 

4

OFF

ZERO

DISPLAY:

 

L-O-H

 

 

 

 

 

 

 

Switch #

Function

Comment

5

6

RANGE

SAMPLE TIME

ON

ON

0

(.000

- .099PPM)

30 SECONDS

ON

OFF

1

(0.10

- 0.99 PPM)

25 SECONDS

OFF

ON

2

(01.0

- 09.9 PPM)

16 SECONDS

OFF

OFF

3

(10 - 50 PPM)

13 SECONDS

AZI Customer Service 800-528-7411 or 602-470-1414

Page 39 of 50

Image 39
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Features Optional AccessoriesIntroduction Principle of Operation ApplicationsZero Air Filter Gas Flow SchematicsAZI Customer Service 800-528-7411 or Instrument Operation LCD CodesDisplay to read H after sampling has started Daily Operations Do not Interrupt this Cycle Sensor RegenerationDo not turn the Zero Adjust potentiometer between samples Zero AdjustAccuracy at MID-RANGE Sample ModeSurvey Mode Operating on AC Power or Generator Operating on Internal Battery PowerExternal battery power Charging Batteries Obtaining Maximum Battery LifePreventive Maintenance Calendar MaintenanceFlow System Part Part NumberInch Fritware Filter Internal Filters Setting the Input Voltage Replacing the Battery PackChanging the Fuse Verification of Calibration and Quality Control CalibrationTroubleshooting Power ProblemsRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Instrument I/O Interface Optional Communications CapabilityAZI Customer Service 800-528-7411 or Functional Test Module Accessories & Maintenance PartsJerome Communication Software Kit Jerome Data LoggerSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Simple Operation Appendix a 631-X Functional Test ModuleAppendix B Jerome Communications Software Instrument Control with Real-Time GraphicsMain JCI Kit Contents System RequirementsData Logger Option Switch # Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Position Function CommentAppendix D Jerome 631-X Option Board Auto-ZeroInstrument Zeroing Timed Regeneration Auto-Sample MA Analog OutputSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Required Parts Suggested Part Similar AZI P/N Fresh Air SolenoidDC Power Adaptor Kit, AZI P/N Y031 DC Power OperationAZI Customer Service 800-528-7411 or Appendix E Calculation of Dynacalibrator H2S Concentration Molecular weight of H 2SWarranty Arizona Instrument LLC Part Number January Trademark and Copyright Protection