Arizona Jerome 631-X operation manual Connection and Setup

Page 43

Jerome 631-X instruments shipped after early 1995, are capable of providing 0-2 volts analog output. Instruments shipped before that time can be upgraded by a firmware update and adjustment.

Instruments that are capable of 0-2 volt output can be upgraded to the 4-20 mA output with the addition of an option board upgrade. This must be installed at the factory.

Connection and Setup:

0-2 volt devices connect as shown in Figure 1. If the instrument includes an option board, be sure its analog jumper (SW101) is set to the “V” position.

The 4-20 mA active receivers connect as shown in Figure 2. The active receiver must contain a voltage source to power the loop current. The receiver must also have an isolated input circuit. That is, it must not be connected to ground or to a voltage source referenced to ground. Be sure that jumper SW101 is set to the “I” position before power is applied.

AZI Customer Service 800-528-7411 or 602-470-1414

Page 43 of 50

Image 43
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Optional Accessories FeaturesIntroduction Principle of Operation ApplicationsZero Air Filter Gas Flow SchematicsAZI Customer Service 800-528-7411 or LCD Codes Instrument OperationDisplay to read H after sampling has started Daily Operations Do not Interrupt this Cycle Sensor RegenerationDo not turn the Zero Adjust potentiometer between samples Zero AdjustAccuracy at MID-RANGE Sample ModeSurvey Mode Operating on Internal Battery Power Operating on AC Power or GeneratorExternal battery power Charging Batteries Obtaining Maximum Battery LifePreventive Maintenance Calendar MaintenancePart Part Number Flow SystemInch Fritware Filter Internal Filters Setting the Input Voltage Replacing the Battery PackChanging the Fuse Verification of Calibration and Quality Control CalibrationPower Problems TroubleshootingRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Instrument I/O Interface Optional Communications CapabilityAZI Customer Service 800-528-7411 or Functional Test Module Accessories & Maintenance PartsJerome Communication Software Kit Jerome Data LoggerSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Simple Operation Appendix a 631-X Functional Test ModuleInstrument Control with Real-Time Graphics Appendix B Jerome Communications SoftwareMain System Requirements JCI Kit ContentsData Logger Option Switch # Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Position Function CommentAuto-Zero Appendix D Jerome 631-X Option BoardInstrument Zeroing Timed Regeneration MA Analog Output Auto-SampleSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Required Parts Suggested Part Similar AZI P/N Fresh Air SolenoidDC Power Adaptor Kit, AZI P/N Y031 DC Power OperationAZI Customer Service 800-528-7411 or Appendix E Calculation of Dynacalibrator H2S Concentration Molecular weight of H 2SWarranty Arizona Instrument LLC Part Number January Trademark and Copyright Protection