Arizona operation manual Jerome 631-X Technical Specifications

Page 27

8. JEROME 631-X TECHNICAL SPECIFICATIONS

Range

0.003ppm (3ppb) to 50ppm H2S in four graduated ranges

Sensitivity

0.003ppm H2S

Precision

5% relative standard deviation

 

 

Accuracy

Range 0: " 0.003ppm at 0.050ppm H2S

 

Range 1: " 0.03ppm at 0.50ppm H2S

 

Range 2: " 0.3ppm at 5.0ppm H2S

 

Range 3: " 2ppm at 25ppm H2S

Response time-sample mode

13 seconds

10 to 50 ppm

1.0 to 10.0 ppm

16 seconds

0.10 to 1.00 ppm

25 seconds

0.001 to 0.100 ppm

30 seconds

Response time-survey mode

3 seconds

0 to 50 ppm

1.0 to 9.9 ppm

6 seconds

0.10 to 0.99 ppm

15 seconds

0.001 to 0.099 ppm

20 seconds

Flow rate

150cc/min (0.15 liters/min)

 

 

Power requirements

100-120 V~, 50/60 Hz, 1 A or 220-240 V~, 50/60 Hz, 1 A

 

 

Fuse

F1A T250V, 5mm X 20mm

 

 

Internal battery pack

Rechargeable Nickel Cadmium

 

 

Operating environment

0E to 40EC, non-condensing, non-explosive

 

 

Case construction

Aluminum alloy

 

 

Dimensions

15 cm x 33 cm x 10 cm (6" w x 13" l x 4" h)

 

 

Weight

3.18 kilos (7 pounds)

 

 

Digital meter display

Liquid crystal display (LCD)

 

 

Certification

CE mark on 220-240 V~, 631-XE model only.

 

 

AZI Customer Service 800-528-7411 or 602-470-1414

Page 27 of 50

Image 27
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Features Optional AccessoriesIntroduction Principle of Operation ApplicationsZero Air Filter Gas Flow SchematicsAZI Customer Service 800-528-7411 or Instrument Operation LCD CodesDisplay to read H after sampling has started Daily Operations Do not Interrupt this Cycle Sensor RegenerationDo not turn the Zero Adjust potentiometer between samples Zero AdjustAccuracy at MID-RANGE Sample ModeSurvey Mode Operating on AC Power or Generator Operating on Internal Battery PowerExternal battery power Charging Batteries Obtaining Maximum Battery LifePreventive Maintenance Calendar MaintenanceFlow System Part Part NumberInch Fritware Filter Internal Filters Setting the Input Voltage Replacing the Battery PackChanging the Fuse Verification of Calibration and Quality Control CalibrationTroubleshooting Power ProblemsRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Instrument I/O Interface Optional Communications CapabilityAZI Customer Service 800-528-7411 or Functional Test Module Accessories & Maintenance PartsJerome Communication Software Kit Jerome Data LoggerSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Simple Operation Appendix a 631-X Functional Test ModuleAppendix B Jerome Communications Software Instrument Control with Real-Time GraphicsMain JCI Kit Contents System RequirementsData Logger Option Switch # Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Position Function CommentAppendix D Jerome 631-X Option Board Auto-ZeroInstrument Zeroing Timed Regeneration Auto-Sample MA Analog OutputSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Required Parts Suggested Part Similar AZI P/N Fresh Air SolenoidDC Power Adaptor Kit, AZI P/N Y031 DC Power OperationAZI Customer Service 800-528-7411 or Appendix E Calculation of Dynacalibrator H2S Concentration Molecular weight of H 2SWarranty Arizona Instrument LLC Part Number January Trademark and Copyright Protection