•In some cases, the instrument cannot resume sampling after regeneration. .L.L.L appears on the display when the ZERO button is pressed and the error message “manual bridge adjust needed” is added to the notes column of the JCS text file when the JCS is used. If this problem persists, it may be necessary to
•When necessary to
Turn the instrument off.
Make a note of the original DIP switch settings of SW100 on the option board. On red DIP switch on the control board, SW2, turn DIP switch 4 to ON.
Set the switches on the option board’s blue DIP box, SW100, to 1,2,6 OFF; 3,4,5 ON. Turn the instrument ON.
Press the Zero button and adjust the potentiometer on top of the instrument until the numbers read between 5 and 7.
Switch option board DIP #1 OFF and ON three times, leaving it ON.
While pressing the ZERO button, turn the potentiometer on the option board until the numbers read between 5 and 7. Note the display will flicker one digit.
Return all switches to their original position.
NOTE: The higher the
Timed Regeneration
If the unit is to be operated unattended for extended periods, AZI recommends that the sensor be regenerated regularly. While operation under JCI or data logger control automatically regenerates saturated sensors. Regeneration will not occur as shown when sample streams have a very low concentration of H2S. The option board controls regeneration on a regular basis, every 6, 12 or 24 hours.
The regeneration intervals are set through a combination of switch settings as shown in the following table:
REGENERATION | ||
Switch #1 | Switch #2 | Interval (Hrs.) |
|
|
|
OFF | OFF | OFF |
|
|
|
ON | OFF | 6 |
|
|
|
OFF | ON | 12 |
|
|
|
ON | ON | 24 |
|
|
|