Page 16

Operating on AC Power or Generator

For stationary use, the 631-X may be operated on AC power.

Operating the instrument on AC power at all times eliminates the need for the battery pack and its necessary maintenance.

The battery may be unplugged or removed completely whenever the instrument is operating on AC power.

When a generator is used to power the Jerome 631-X, a high quality line conditioner or voltage regulator is required to ensure a pure sine wave and regulated voltage is applied to the instrument. The gold film sensor may be damaged by voltage that varies in amplitude or by surges, spikes, and/or noise on the power line.

Operating on Internal Battery Power

For portable use, the 631-X may be operated on Battery power.

When you operate the instrument on battery power, please be aware of the following:

A fully charged battery pack, AZI P/N Z4000-0907, provides power for a minimum of six (6) hours of operation.

For operating more than six (6) hours, an extra fully charged battery pack is needed. Complete battery recharging takes 14 hours. Refer to Charging Batteries on page 17. The 631-X uses a rechargeable Nickel Cadmium (NiCad ) battery. Dispose of worn- out batteries properly when you are replacing the battery pack.

External battery power

A special version of the Jerome 631-X and a DC Power Kit are available to operate the instrument from a secondary DC source. The source may be a car/truck battery or a storage cell used in conjunction with solar panels.

Call AZI Customer Service at 800-528-7411, 602-470-1414, or e-mail support@azic.com for additional information.

Page 16 of 50

AZI Customer Service 800-528-7411 or 602-470-1414

Image 16
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Optional Accessories FeaturesIntroduction Applications Principle of OperationGas Flow Schematics Zero Air FilterAZI Customer Service 800-528-7411 or LCD Codes Instrument OperationDisplay to read H after sampling has started Daily Operations Sensor Regeneration Do not Interrupt this CycleZero Adjust Do not turn the Zero Adjust potentiometer between samplesSample Mode Accuracy at MID-RANGESurvey Mode Operating on Internal Battery Power Operating on AC Power or GeneratorExternal battery power Obtaining Maximum Battery Life Charging BatteriesMaintenance Preventive Maintenance CalendarPart Part Number Flow SystemInch Fritware Filter Internal Filters Replacing the Battery Pack Setting the Input VoltageChanging the Fuse Calibration Verification of Calibration and Quality ControlPower Problems TroubleshootingRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Optional Communications Capability Instrument I/O InterfaceAZI Customer Service 800-528-7411 or Accessories & Maintenance Parts Functional Test ModuleJerome Data Logger Jerome Communication Software KitSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Appendix a 631-X Functional Test Module Simple OperationInstrument Control with Real-Time Graphics Appendix B Jerome Communications SoftwareMain System Requirements JCI Kit ContentsData Logger Option Appendix C Internal DIP Switch Settings DIP Switch SettingsSwitch # Position Function Comment Switch # Function CommentAuto-Zero Appendix D Jerome 631-X Option BoardInstrument Zeroing Timed Regeneration MA Analog Output Auto-SampleSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Fresh Air Solenoid Required Parts Suggested Part Similar AZI P/NDC Power Operation DC Power Adaptor Kit, AZI P/N Y031AZI Customer Service 800-528-7411 or Molecular weight of H 2S Appendix E Calculation of Dynacalibrator H2S ConcentrationWarranty Trademark and Copyright Protection Arizona Instrument LLC Part Number January