Miscellaneous Problems

Display reads .P.P.P when

Power cord not attached.

Check power cord for connection

regeneration is attempted.

Blown fuse.

Replace fuse. Refer to page 22.

 

 

Line voltage less than 100

Check line voltage settings. Refer

 

VAC (or less than 200 VAC

to page 21.

 

for 220 unit).

 

 

Cycles dipswitch set

Check input cycle settings. Refer to

 

incorrectly.

page 39.

 

 

If fuse and line voltage are OK, it

 

 

may be circuit board adjustment or

 

 

component failure. Call AZI

 

 

Customer Service at 800-528-7411

 

 

or 602 470-1414.

Display reads .E.E.E

Very low battery.

Recharge battery. Refer to page 17.

 

 

 

 

Replace battery. Refer to page 21.

Page 26 of 50

AZI Customer Service 800-528-7411 or 602-470-1414

Image 26
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Introduction FeaturesOptional Accessories Applications Principle of OperationGas Flow Schematics Zero Air FilterAZI Customer Service 800-528-7411 or Display to read H after sampling has started Instrument OperationLCD Codes Daily Operations Sensor Regeneration Do not Interrupt this CycleZero Adjust Do not turn the Zero Adjust potentiometer between samplesSample Mode Accuracy at MID-RANGESurvey Mode External battery power Operating on AC Power or GeneratorOperating on Internal Battery Power Obtaining Maximum Battery Life Charging BatteriesMaintenance Preventive Maintenance CalendarInch Fritware Filter Flow SystemPart Part Number Internal Filters Replacing the Battery Pack Setting the Input VoltageChanging the Fuse Calibration Verification of Calibration and Quality ControlRegeneration & Zero Problems TroubleshootingPower Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Optional Communications Capability Instrument I/O InterfaceAZI Customer Service 800-528-7411 or Accessories & Maintenance Parts Functional Test ModuleJerome Data Logger Jerome Communication Software KitSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Appendix a 631-X Functional Test Module Simple OperationMain Appendix B Jerome Communications SoftwareInstrument Control with Real-Time Graphics Data Logger Option JCI Kit ContentsSystem Requirements Switch # Position Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Function CommentInstrument Zeroing Appendix D Jerome 631-X Option BoardAuto-Zero Timed Regeneration SW101 Functions Auto-SampleMA Analog Output Connection and Setup AZI Customer Service 800-528-7411 or Fresh Air Solenoid Required Parts Suggested Part Similar AZI P/NDC Power Operation DC Power Adaptor Kit, AZI P/N Y031AZI Customer Service 800-528-7411 or Molecular weight of H 2S Appendix E Calculation of Dynacalibrator H2S ConcentrationWarranty Trademark and Copyright Protection Arizona Instrument LLC Part Number January