Arizona Jerome 631-X operation manual Survey Mode

Page 15

Survey Mode

The survey mode takes samples every 3 to 20 seconds automatically. The length of time varies with the hydrogen sulfide concentration. Use this mode to locate the source of hydrogen sulfide, such as a leak, a hot spot, or to assess areas of potentially high hydrogen sulfide concentrations. After the survey mode is activated, the 631-X samples continuously.

Press the power ON button.

The digital meter displays 000.

Disregard the LCD’s initial momentary readings.

Recharge or replace the battery pack if the LO BAT indicator REMAINS ON, see page 17 and/or page 21 if necessary.

Allow a 1-minute warm up before beginning the next step to ensure the instrument's electronics have stabilized.

Lock the instrument in the survey mode:

Hold the SAMPLE button down until the sensor status indicator bar(s) begin flashing on the display.

Press the ZERO button, then release the SAMPLE button.

The pump should continue to run and the display should update every survey cycle.

The instrument remains in the survey mode until one of the following occurs: The sensor is saturated.

A LO BAT (low battery) signal is encountered.

An HL (high hydrogen sulfide level) is encountered. The instrument is turned OFF.

Press the power OFF button to end the survey mode.

RANGE

CONCENTRATION

RESPONSE TIME

0

0.001 to 0 .099 ppm

20 Seconds

1

0.10 to 0.99 ppm

15 Seconds

2

1.0 to 9.9 ppm

6 Seconds

3

10 to 50 ppm

3 Seconds

NOTE: Approximately 100 samples at 0.5 ppm may be taken before the sensor saturates and regeneration is required.

AZI Customer Service 800-528-7411 or 602-470-1414

Page 15 of 50

Image 15
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Features Optional AccessoriesIntroduction Principle of Operation ApplicationsZero Air Filter Gas Flow SchematicsAZI Customer Service 800-528-7411 or Instrument Operation LCD CodesDisplay to read H after sampling has started Daily Operations Do not Interrupt this Cycle Sensor RegenerationDo not turn the Zero Adjust potentiometer between samples Zero AdjustAccuracy at MID-RANGE Sample ModeSurvey Mode Operating on AC Power or Generator Operating on Internal Battery PowerExternal battery power Charging Batteries Obtaining Maximum Battery LifePreventive Maintenance Calendar MaintenanceFlow System Part Part NumberInch Fritware Filter Internal Filters Setting the Input Voltage Replacing the Battery PackChanging the Fuse Verification of Calibration and Quality Control CalibrationTroubleshooting Power ProblemsRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Instrument I/O Interface Optional Communications CapabilityAZI Customer Service 800-528-7411 or Functional Test Module Accessories & Maintenance PartsJerome Communication Software Kit Jerome Data LoggerSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Simple Operation Appendix a 631-X Functional Test ModuleAppendix B Jerome Communications Software Instrument Control with Real-Time GraphicsMain JCI Kit Contents System RequirementsData Logger Option Switch # Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Position Function CommentAppendix D Jerome 631-X Option Board Auto-ZeroInstrument Zeroing Timed Regeneration Auto-Sample MA Analog OutputSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Required Parts Suggested Part Similar AZI P/N Fresh Air SolenoidDC Power Adaptor Kit, AZI P/N Y031 DC Power OperationAZI Customer Service 800-528-7411 or Appendix E Calculation of Dynacalibrator H2S Concentration Molecular weight of H 2SWarranty Arizona Instrument LLC Part Number January Trademark and Copyright Protection