Arizona Jerome 631-X operation manual Daily Operations

Page 11

Daily Operations

Before each day's use of the Jerome 631-X, perform the following steps to verify proper instrument operation:

Press the power ON button.

The digital meter displays 000.

(Disregard the digital meter's initial momentary reading.)

Recharge or replace the battery pack if the LO BAT indicator REMAINS ON. Refer to “Charging Batteries” on page 17 and/or “Replacing the Battery Pack” on page 21.

To ensure the instrument's electronics have stabilized, allow a 1-minute warm up before beginning the next step.

Use the Zero Air Filter to equilibrate the instrument to ambient air temperature. Install the Zero Air Filter in the instrument’s intake.

Sample continuously until the readings stabilize.

Perform sensor regeneration. Refer to page 12 for the procedure.

Thirty minutes after sensor regeneration is complete, zero the instrument. See page 13.

NOTE: For maximum accuracy, such as when testing with the Functional Test Module, wait 30 minutes after the sensor regeneration cycle to re-zero the unit. For immediate use, the unit can be re-zeroed immediately after sensor regeneration. See the notes on page 13.

• Press the SAMPLE button.

During the sample cycle, the digital meter displays bars (-, --, or ---) to indicate the amount of sensor saturation.

At the end of the sampling cycle, read the digital

meter.

The number shown on the digital meter is the hydrogen sulfide concentration in ppm.

This value remains on the display until the next sample is taken.

The digital meter automatically zeroes at the start of each sample.

• At the end of each day's use, perform sensor regeneration as described in the next section.

DO NOT ALLOW HYDROGEN SULFIDE TO STAY

ON THE GOLD FILM SENSOR OVERNIGHT.

AZI Customer Service 800-528-7411 or 602-470-1414

Page 11 of 50

Image 11
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Introduction FeaturesOptional Accessories Principle of Operation ApplicationsZero Air Filter Gas Flow SchematicsAZI Customer Service 800-528-7411 or Display to read H after sampling has started Instrument OperationLCD Codes Daily Operations Do not Interrupt this Cycle Sensor RegenerationDo not turn the Zero Adjust potentiometer between samples Zero AdjustAccuracy at MID-RANGE Sample ModeSurvey Mode External battery power Operating on AC Power or GeneratorOperating on Internal Battery Power Charging Batteries Obtaining Maximum Battery LifePreventive Maintenance Calendar MaintenanceInch Fritware Filter Flow SystemPart Part Number Internal Filters Setting the Input Voltage Replacing the Battery PackChanging the Fuse Verification of Calibration and Quality Control CalibrationRegeneration & Zero Problems TroubleshootingPower Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Instrument I/O Interface Optional Communications CapabilityAZI Customer Service 800-528-7411 or Functional Test Module Accessories & Maintenance PartsJerome Communication Software Kit Jerome Data LoggerSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Simple Operation Appendix a 631-X Functional Test ModuleMain Appendix B Jerome Communications SoftwareInstrument Control with Real-Time Graphics Data Logger Option JCI Kit ContentsSystem Requirements Switch # Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Position Function CommentInstrument Zeroing Appendix D Jerome 631-X Option BoardAuto-Zero Timed Regeneration SW101 Functions Auto-SampleMA Analog Output Connection and Setup AZI Customer Service 800-528-7411 or Required Parts Suggested Part Similar AZI P/N Fresh Air SolenoidDC Power Adaptor Kit, AZI P/N Y031 DC Power OperationAZI Customer Service 800-528-7411 or Appendix E Calculation of Dynacalibrator H2S Concentration Molecular weight of H 2SWarranty Arizona Instrument LLC Part Number January Trademark and Copyright Protection