Arizona Jerome 631-X operation manual Changing the Fuse

Page 22

Changing the Fuse

If the instrument display reads .P.P.P when the instrument is connected to AC power or when REGEN is pressed, or if the battery will not charge, the fuse may need to be replaced. The AC line power could also be less than 100 VAC (220 VAC). Check the fuse with an ohmmeter and the AC line power with a voltage meter.

Locate the power receptacle on the rear of the instrument.

Insert a small screwdriver in the slot, located in the power receptacle, and gently slide the fuse compartment out.

If the fuse in the open-sided clip is open, remove and discard it.

Replace the discarded fuse with the spare fuse located in the slide-out spare fuse compartment.

Replace the fuse compartment in the power receptacle.

As soon as possible, replace the spare fuse with another 1A, 250V, time delay fuse, AZI P/N 5100 1012).

Page 22 of 50

AZI Customer Service 800-528-7411 or 602-470-1414

Image 22
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Optional Accessories FeaturesIntroduction Applications Principle of OperationGas Flow Schematics Zero Air FilterAZI Customer Service 800-528-7411 or LCD Codes Instrument OperationDisplay to read H after sampling has started Daily Operations Sensor Regeneration Do not Interrupt this CycleZero Adjust Do not turn the Zero Adjust potentiometer between samplesSample Mode Accuracy at MID-RANGESurvey Mode Operating on Internal Battery Power Operating on AC Power or GeneratorExternal battery power Obtaining Maximum Battery Life Charging BatteriesMaintenance Preventive Maintenance CalendarPart Part Number Flow SystemInch Fritware Filter Internal Filters Replacing the Battery Pack Setting the Input VoltageChanging the Fuse Calibration Verification of Calibration and Quality ControlPower Problems TroubleshootingRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Optional Communications Capability Instrument I/O InterfaceAZI Customer Service 800-528-7411 or Accessories & Maintenance Parts Functional Test ModuleJerome Data Logger Jerome Communication Software KitSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Appendix a 631-X Functional Test Module Simple OperationInstrument Control with Real-Time Graphics Appendix B Jerome Communications SoftwareMain System Requirements JCI Kit ContentsData Logger Option Switch # Position Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Function CommentAuto-Zero Appendix D Jerome 631-X Option BoardInstrument Zeroing Timed Regeneration MA Analog Output Auto-SampleSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Fresh Air Solenoid Required Parts Suggested Part Similar AZI P/NDC Power Operation DC Power Adaptor Kit, AZI P/N Y031AZI Customer Service 800-528-7411 or Molecular weight of H 2S Appendix E Calculation of Dynacalibrator H2S ConcentrationWarranty Trademark and Copyright Protection Arizona Instrument LLC Part Number January