Arizona Jerome 631-X operation manual Flow System, Part Part Number, Inch Fritware Filter

Page 19

Flow System

The Jerome 631-X's flow system is the crucial link between the sensor and the sample. For the instrument to perform correctly, the flow system must be properly maintained. The user maintainable components of this system are the intake filter (.25 inch fritware), two scrubber filters and connecting tubing.

Check the Preventive Maintenance Calendar on page 18, for a suggested schedule for changing fritware and scrubber filters. The Tygon™ tubing in the system must be free of crimps for proper flow.

Part

Part Number

Scrubber Filter

Z2600 3930

LFS Scrubber Filter

Z2600 3933

LFD Scrubber Filter

Z2600 3934

.25 inch Fritware Filter

2600 3039

Tygon™ Tubing - 1/8" I.D. (1')

2500 3001

Tygon™ Tubing - 1/16" I.D. (1')

2500 3002

.25 inch Fritware Filter

Replace the .25 inch fritware filter once each week or as needed. In dusty environments, the fritware filter may need to be replaced as often as once a day. Replacement .25-inch fritware filters are available from AZI, Consumable Sales at 800-528-7411 or 602-470-1414.

Unscrew and remove the intake.

Push the old fritware filter disc out of the intake with your trimmer tool.

Avoid touching the new fritware disc with fingers. Use tweezers to insert the new fritware.

Use the blunt end of the trimmer tool to seat the fritware disc firmly against the inner ledge of the intake.

Screw the intake back on the Jerome 631-X.

CAUTION:

The stem coming from the instrument onto which the outer intake housing is attached must be securely held in place. If loose, the tubing inside the instrument can become twisted when the intake housing is replaced. It may be necessary to open the instrument and tighten the hold-down nuts inside the instrument. Call AZI Customer Service at 800-528-7411, 601-470-1414, or e-mail support@azic.com if you have any questions

AZI Customer Service 800-528-7411 or 602-470-1414

Page 19 of 50

Image 19
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Optional Accessories FeaturesIntroduction Principle of Operation ApplicationsZero Air Filter Gas Flow SchematicsAZI Customer Service 800-528-7411 or LCD Codes Instrument OperationDisplay to read H after sampling has started Daily Operations Do not Interrupt this Cycle Sensor RegenerationDo not turn the Zero Adjust potentiometer between samples Zero AdjustAccuracy at MID-RANGE Sample ModeSurvey Mode Operating on Internal Battery Power Operating on AC Power or GeneratorExternal battery power Charging Batteries Obtaining Maximum Battery LifePreventive Maintenance Calendar MaintenancePart Part Number Flow SystemInch Fritware Filter Internal Filters Setting the Input Voltage Replacing the Battery PackChanging the Fuse Verification of Calibration and Quality Control CalibrationPower Problems TroubleshootingRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Instrument I/O Interface Optional Communications CapabilityAZI Customer Service 800-528-7411 or Functional Test Module Accessories & Maintenance PartsJerome Communication Software Kit Jerome Data LoggerSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Simple Operation Appendix a 631-X Functional Test ModuleInstrument Control with Real-Time Graphics Appendix B Jerome Communications SoftwareMain System Requirements JCI Kit ContentsData Logger Option Switch # Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Position Function CommentAuto-Zero Appendix D Jerome 631-X Option BoardInstrument Zeroing Timed Regeneration MA Analog Output Auto-SampleSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Required Parts Suggested Part Similar AZI P/N Fresh Air SolenoidDC Power Adaptor Kit, AZI P/N Y031 DC Power OperationAZI Customer Service 800-528-7411 or Appendix E Calculation of Dynacalibrator H2S Concentration Molecular weight of H 2SWarranty Arizona Instrument LLC Part Number January Trademark and Copyright Protection