Arizona Jerome 631-X operation manual Replacing the Battery Pack, Setting the Input Voltage

Page 21

Replacing the Battery Pack

Press the power OFF button.

Unplug the power cord.

Remove the two (2) side screws from the intake end of the instrument and open the case lid.

Disconnect the battery connector from the board.

Loosen the two (2) captive screws holding the battery bracket and remove the bracket.

Remove the old battery pack and replace with a new battery pack.

Replace the battery bracket and tighten the captive screws.

Connect the new battery connector to the board.

Close the case and replace the two (2) side screws.

Dispose of the old NiCad battery in accordance with state and federal regulations.

Setting the Input Voltage

Instruments are factory set and calibrated to use the power setting requested on the order. However, the voltage setting is easily changed to use either 110 or 220 VAC.

Ensure the instrument is turned OFF and unplugged.

Locate the voltage selector on the rear of the instrument.

Insert a small screwdriver in the voltage selector slot and turn the selector until the arrow points toward your setting choice and a click is heard.

AZI Customer Service 800-528-7411 or 602-470-1414

Page 21 of 50

Image 21
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Features Optional AccessoriesIntroduction Principle of Operation ApplicationsZero Air Filter Gas Flow SchematicsAZI Customer Service 800-528-7411 or Instrument Operation LCD CodesDisplay to read H after sampling has started Daily Operations Do not Interrupt this Cycle Sensor RegenerationDo not turn the Zero Adjust potentiometer between samples Zero AdjustAccuracy at MID-RANGE Sample ModeSurvey Mode Operating on AC Power or Generator Operating on Internal Battery PowerExternal battery power Charging Batteries Obtaining Maximum Battery LifePreventive Maintenance Calendar MaintenanceFlow System Part Part NumberInch Fritware Filter Internal Filters Setting the Input Voltage Replacing the Battery PackChanging the Fuse Verification of Calibration and Quality Control CalibrationTroubleshooting Power ProblemsRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Instrument I/O Interface Optional Communications CapabilityAZI Customer Service 800-528-7411 or Functional Test Module Accessories & Maintenance PartsJerome Communication Software Kit Jerome Data LoggerSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Simple Operation Appendix a 631-X Functional Test ModuleAppendix B Jerome Communications Software Instrument Control with Real-Time GraphicsMain JCI Kit Contents System RequirementsData Logger Option DIP Switch Settings Appendix C Internal DIP Switch SettingsSwitch # Position Function Comment Switch # Function CommentAppendix D Jerome 631-X Option Board Auto-ZeroInstrument Zeroing Timed Regeneration Auto-Sample MA Analog OutputSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Required Parts Suggested Part Similar AZI P/N Fresh Air SolenoidDC Power Adaptor Kit, AZI P/N Y031 DC Power OperationAZI Customer Service 800-528-7411 or Appendix E Calculation of Dynacalibrator H2S Concentration Molecular weight of H 2SWarranty Arizona Instrument LLC Part Number January Trademark and Copyright Protection