Arizona Jerome 631-X operation manual Auto-Sample, MA Analog Output, SW101 Functions

Page 42

Auto-Sample

If a data logger is connected and operating in the manual sampling mode or a data logger is not connected the following automatic sampling rates may be selected with SW100 dip switch settings:

 

Dip switch settings

Sampling frequency

 

 

 

 

3

 

4

5

 

 

ON

 

ON

ON

No automatic sampling

 

 

 

 

 

OFF

 

ON

ON

5 minutes

 

 

 

 

 

OFF

 

OFF

ON

15 minutes

OFF

 

ON

OFF

30 minutes

 

 

 

 

 

OFF

 

OFF

OFF

1 hour

 

 

 

 

 

This auto-sample function will not function if a data logger is connected and operating in automatic sampler mode programmed through the JCS.

4-20 MA Analog Output

The analog output signal at pin 18 of the 25 pin connector can be configured to provide the instrument's native mode 0-2 Volt output or the optional 4-20 mA output by setting the option board jumper (SW101) to the "V" position for voltage, or the "I" position for current. (Pin 23 is the ground pin for the analog output function. Pin 18 is positive with respect to the ground pin).

The 0-2 Volt output circuit can drive loads of 10 k ohms or higher.

The 4-20 mA output is a passive transmitter and requires the connected receiver to supply between 10 and 28 volts of excitation potential.

The 631-X must have the auto ranging feature disabled in order to get accurate analog output readings. The approximate range of the H2S concentration must be known before the instrument can be set into that specific range (0,1, 2 or 3). The analog output signal is based on each individual range and not the entire range (0-50 ppm).

Note that neither analog output circuit is floating. The negative terminals of both circuits are connected to the instrument's common ground buss.

SW101 Functions:

V = 0-2V analog output

I = 4-20 ma analog output

Page 42 of 50

AZI Customer Service 800-528-7411 or 602-470-1414

Image 42
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Features Optional AccessoriesIntroduction Applications Principle of OperationGas Flow Schematics Zero Air FilterAZI Customer Service 800-528-7411 or Instrument Operation LCD CodesDisplay to read H after sampling has started Daily Operations Sensor Regeneration Do not Interrupt this CycleZero Adjust Do not turn the Zero Adjust potentiometer between samplesSample Mode Accuracy at MID-RANGESurvey Mode Operating on AC Power or Generator Operating on Internal Battery PowerExternal battery power Obtaining Maximum Battery Life Charging BatteriesMaintenance Preventive Maintenance CalendarFlow System Part Part NumberInch Fritware Filter Internal Filters Replacing the Battery Pack Setting the Input VoltageChanging the Fuse Calibration Verification of Calibration and Quality ControlTroubleshooting Power ProblemsRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Optional Communications Capability Instrument I/O InterfaceAZI Customer Service 800-528-7411 or Accessories & Maintenance Parts Functional Test ModuleJerome Data Logger Jerome Communication Software KitSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Appendix a 631-X Functional Test Module Simple OperationAppendix B Jerome Communications Software Instrument Control with Real-Time GraphicsMain JCI Kit Contents System RequirementsData Logger Option Switch # Position Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Function CommentAppendix D Jerome 631-X Option Board Auto-ZeroInstrument Zeroing Timed Regeneration Auto-Sample MA Analog OutputSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Fresh Air Solenoid Required Parts Suggested Part Similar AZI P/NDC Power Operation DC Power Adaptor Kit, AZI P/N Y031AZI Customer Service 800-528-7411 or Molecular weight of H 2S Appendix E Calculation of Dynacalibrator H2S ConcentrationWarranty Trademark and Copyright Protection Arizona Instrument LLC Part Number January