Arizona Jerome 631-X operation manual Internal Filters

Page 20

Internal Filters

Replace the internal filters after six (6) months of use, or as needed.

Press the power OFF button and unplug the power cord.

Remove the two (2) side screws from the intake end of the instrument and open the case.

Carefully disconnect the Tygon tubing from both ends of the filters and discard the old filters.

CAUTION:

Scrubber filters contain Resisorb. Used filters, especially the scrubber filter may contain hydrogen sulfide also. Use proper methods when disposing of used filters. Call AZI Customer Service at 800-528-7411, 601-470-1414, or e-mail support@azic.com for a copy of the Resisorb MSDS or for other questions.

Connect the new filters to the Tygon tubing, ensuring all straight hose barbs point toward the intake/pump corner of the case and elbow hose barbs point toward the sensor housing as shown in the illustration.

Push the Tygon as far as it will go onto the filter fittings.

Push the filters into the mounting clips.

Remove any crimps or twists in the tubing and ensure that tubing connections are secure. If the tubing is loose, readings may not be accurate. Replace any tubing that has deteriorated due to heat and/or age.

Close the case and replace the screws.

Dispose of all filters in accordance with state and federal environmental regulations.

Page 20 of 50

AZI Customer Service 800-528-7411 or 602-470-1414

Image 20
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Introduction FeaturesOptional Accessories Applications Principle of OperationGas Flow Schematics Zero Air FilterAZI Customer Service 800-528-7411 or Display to read H after sampling has started Instrument OperationLCD Codes Daily Operations Sensor Regeneration Do not Interrupt this CycleZero Adjust Do not turn the Zero Adjust potentiometer between samplesSample Mode Accuracy at MID-RANGESurvey Mode External battery power Operating on AC Power or GeneratorOperating on Internal Battery Power Obtaining Maximum Battery Life Charging BatteriesMaintenance Preventive Maintenance CalendarInch Fritware Filter Flow SystemPart Part Number Internal Filters Replacing the Battery Pack Setting the Input VoltageChanging the Fuse Calibration Verification of Calibration and Quality ControlRegeneration & Zero Problems TroubleshootingPower Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Optional Communications Capability Instrument I/O InterfaceAZI Customer Service 800-528-7411 or Accessories & Maintenance Parts Functional Test ModuleJerome Data Logger Jerome Communication Software KitSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Appendix a 631-X Functional Test Module Simple OperationMain Appendix B Jerome Communications SoftwareInstrument Control with Real-Time Graphics Data Logger Option JCI Kit ContentsSystem Requirements Appendix C Internal DIP Switch Settings DIP Switch SettingsSwitch # Position Function Comment Switch # Function CommentInstrument Zeroing Appendix D Jerome 631-X Option BoardAuto-Zero Timed Regeneration SW101 Functions Auto-SampleMA Analog Output Connection and Setup AZI Customer Service 800-528-7411 or Fresh Air Solenoid Required Parts Suggested Part Similar AZI P/NDC Power Operation DC Power Adaptor Kit, AZI P/N Y031AZI Customer Service 800-528-7411 or Molecular weight of H 2S Appendix E Calculation of Dynacalibrator H2S ConcentrationWarranty Trademark and Copyright Protection Arizona Instrument LLC Part Number January