Internal Filters
•Replace the internal filters after six (6) months of use, or as needed.
•Press the power OFF button and unplug the power cord.
•Remove the two (2) side screws from the intake end of the instrument and open the case.
•Carefully disconnect the Tygon tubing from both ends of the filters and discard the old filters.
CAUTION:
Scrubber filters contain Resisorb. Used filters, especially the scrubber filter may contain hydrogen sulfide also. Use proper methods when disposing of used filters. Call AZI Customer Service at
•Connect the new filters to the Tygon tubing, ensuring all straight hose barbs point toward the intake/pump corner of the case and elbow hose barbs point toward the sensor housing as shown in the illustration.
Push the Tygon as far as it will go onto the filter fittings.
•Push the filters into the mounting clips.
•Remove any crimps or twists in the tubing and ensure that tubing connections are secure. If the tubing is loose, readings may not be accurate. Replace any tubing that has deteriorated due to heat and/or age.
•Close the case and replace the screws.
•Dispose of all filters in accordance with state and federal environmental regulations.
Page 20 of 50 | AZI Customer Service |