2. INTRODUCTION

The Jerome 631-X Hydrogen Sulfide Analyzer is an ambient air analyzer with a range of 0.003 ppm to 50 ppm (parts per million).

CAUTION:

The Jerome 631-X is for vapor use only.

DO NOT allow the probe or the instrument's intake

to be exposed to any liquid, dust

or other foreign material.

The 631-X is designed to be easy to operate for quick and accurate analysis of hydrogen sulfide vapor levels. It has few maintenance requirements. However, please take a moment to read this manual before attempting operation. If you have any questions about your application or operation, please call AZI Customer Service at (800) 528-7411 or (602) 470-1414 or e-mail support@azic.com for assistance.

631-X Features

Accurate analysis of hydrogen sulfide in seconds

Wide detection range allows multiple applications

Survey mode for rapid source detection of hydrogen sulfide concentrations

Rechargeable internal battery pack for portability

Automatic backlight for LCD during low light conditions

Microprocessor ensures a linear response throughout the entire range of the sensor

Inherently stable gold film sensor

Optional Accessories

Data Logger, P/N Y6100 0057, to record field monitoring information

Jerome Communication Software Kit, P/N Y6100 0054, for unattended fixed-point sampling and downloading information from the data logger to a computer.

Option Board (factory installed option) for external fresh air solenoid support, auto-zeroing, DC power operation, timed regeneration, 4-20 mA or 0-2 V analog output, and timed sampling

Functional Test Module (FTM) and Accessory Kit, P/N Z2600 0918 and Y2600 0902, for sensor calibration verification in the field

Field Carrying Cases, hard sided P/N Y411 0904 or soft P/N 1400 0052, for versatile handling and additional storage

Maintenance Kit, P/N Y631 0905 for routine maintenance and upkeep

Page 6 of 50

AZI Customer Service 800-528-7411 or 602-470-1414

Image 6
Contents Jerome Jerome Table of Contents For Those WHO Cant Read the Whole Manual NOW AZI Customer Service 800-528-7411 or Features Optional AccessoriesIntroduction Applications Principle of OperationGas Flow Schematics Zero Air FilterAZI Customer Service 800-528-7411 or Instrument Operation LCD CodesDisplay to read H after sampling has started Daily Operations Sensor Regeneration Do not Interrupt this CycleZero Adjust Do not turn the Zero Adjust potentiometer between samplesSample Mode Accuracy at MID-RANGESurvey Mode Operating on AC Power or Generator Operating on Internal Battery PowerExternal battery power Obtaining Maximum Battery Life Charging BatteriesMaintenance Preventive Maintenance CalendarFlow System Part Part NumberInch Fritware Filter Internal Filters Replacing the Battery Pack Setting the Input VoltageChanging the Fuse Calibration Verification of Calibration and Quality ControlTroubleshooting Power ProblemsRegeneration & Zero Problems Sampling Problems Miscellaneous Problems Jerome 631-X Technical Specifications Optional Communications Capability Instrument I/O InterfaceAZI Customer Service 800-528-7411 or Accessories & Maintenance Parts Functional Test ModuleJerome Data Logger Jerome Communication Software KitSpare Parts AZI Customer Service 800-528-7411 or AZI Customer Service 800-528-7411 or Factory Calibration Service Appendix a 631-X Functional Test Module Simple OperationAppendix B Jerome Communications Software Instrument Control with Real-Time GraphicsMain JCI Kit Contents System RequirementsData Logger Option Switch # Position Function Comment Appendix C Internal DIP Switch SettingsDIP Switch Settings Switch # Function CommentAppendix D Jerome 631-X Option Board Auto-ZeroInstrument Zeroing Timed Regeneration Auto-Sample MA Analog OutputSW101 Functions Connection and Setup AZI Customer Service 800-528-7411 or Fresh Air Solenoid Required Parts Suggested Part Similar AZI P/NDC Power Operation DC Power Adaptor Kit, AZI P/N Y031AZI Customer Service 800-528-7411 or Molecular weight of H 2S Appendix E Calculation of Dynacalibrator H2S ConcentrationWarranty Trademark and Copyright Protection Arizona Instrument LLC Part Number January