18

Calibrate

Performing a Calibration

6.In the calibration screen, either press the CAL softkey or press the key.

Note: the calibration screen details the last calibration and is saved into the memory of the instrument. If you download your data to DMS (software), the calibration history is saved to the session and is viewable/printable in the software.

Opening CAL screen (to perform calibration)

Either press the key or press the Cal softkey

Figure 4-3: Calibration screen

7. In the calibration screen, press the corresponding softkeys or press / to adjust to the specific dB level (noted on the calibrator).

Calibrating

Match the dB level from the calibrator to the SE- 400 Series by pressing either the corresponding softkeys or the up/down arrow keys

 

Figure 4-4: CAL screen

8.

Press the

key to save the changes or escape to abort.

9.

Press the

key twice to return to the main measurement screen.

Sound Examiner SE-400 Series

Page 25
Image 25
3M SE-400 user manual Calibration screen