74
Data Addendum
Pressure to random incidence corrections with BK4936 microphone, remote preamp, and RICR (section 5.2.7)
Note: added to the pressure response to calculate random incidence response.
| BK4936 microphone , |
| remote preamp, & RICR |
| Random Incidence |
Frequency | Corrections |
in Hz | in dB |
32 | 0.5 |
40 | 0.5 |
50 | 0.4 |
63 | 0.3 |
80 | 0.2 |
100 | 0.1 |
125 | 0.2 |
160 | 0.1 |
200 | 0.1 |
250 | 0.3 |
315 | 0.2 |
400 | 0.3 |
500 | 0.1 |
630 | 0.2 |
800 | 0.0 |
1000 | 0.0 |
1250 | 0.3 |
1600 | 0.1 |
2000 | 0.2 |
2500 | 0.4 |
3150 | 0.7 |
4000 | 1.1 |
5000 | 1.8 |
6300 | 2.7 |
8000 | 3.8 |
10000 | 5.6 |
12500 | 6.1 |
16000 | 5.3 |
Sound Examiner