40
Data Addendum
Data Addendum: SE-400 Series measuring to IEC 61672-1
About this data addendum
This data addendum presents submission information for testing the 3MTM Sound Examiner
Note: The nominal microphone sensitivity is
Linear operating range (sections: 5.5.10, 5.5.11, 9.2.7e, 9.3e, 9.3f)
The starting level for measuring level linearity is 114 dB at all frequencies. Calibration point is 114 dB at 400 mV @ 1 kHz.
Type/Class 1 and Type/Class 2
|
|
|
|
| Equivalent input |
|
|
| voltage | 0.400 VRMS |
|
| Capacitance | 12 pF |
|
| Resistance | >1G |
|
Linear operating range with SPL
Note: Reference level 114 dB, Cal Point: 400mVrms = 114 dB @ 1 kHz
Frequency | SPL |
| LEQ |
|
| Linear Operating Range (dB) | Linear Operating Range (dB) | ||
31.5 | 75 | 140 | 75 | 140 |
1000 | 34 | 140 | 34 | 140 |
4000 | 33 | 140 | 33 | 140 |
8000 | 37 | 140 | 37 | 140 |
12500 | 42 | 140 | 42 | 140 |
Linear operating range with SPL
Note: Reference level 114 dB, Cal Point: 400mVrms = 114 dB @ 1 kHz
Frequency |
| LEQ |
| |
| Linear Operating Range (dB) | Linear Operating Range (dB) | ||
31.5 | 43 | 140 | 43 | 140 |
1000 | 38 | 140 | 38 | 140 |
4000 | 39 | 140 | 39 | 140 |
8000 | 42 | 140 | 42 | 140 |
12500 | 47 | 140 | 47 | 140 |
Sound Examiner