80

Data Addendum

Cable and radio frequency emission (sections: 5.18.1, 5.18.2, 9.3n)

The charging cable is a shielded 1 meter long USB cable.

Battery voltage range and power supply (sections: 5.20.2, 5.20.3, 9.3j)

The charging circuit stops charging when the battery voltage reaches its maximum of 4.1 Volts DC. The unit shuts down when the minimum battery voltage of 3.4 volts DC is reached. The instrument may be operated with a USB power cable.

Electrostatic discharges (section 6.5.2, 9.2.7 b)

Exposure to electrostatic discharges will not change operating state, change of configuration or corruption or loss of stored data.

AC power and radio frequency (sections: 6.6.1, 6.6.3, 6.6.4, 9.2.7c, 9.3o)

No effect was observed in any orientation or configuration of the SE-400 SERIES within a 60 Hz 80 A/m magnetic field. During radio frequency immunity testing, a 3-meter USB cable was connected between the SE-400 Series and a computer USB port. The SE-400 Series was set to display SPL.

Directional windscreen corrections (section: 7.2)

Note: in dB re:20 uPA

 

0 Degree

30 Degree

60 Degree

90 Degree

120 Degree

150 Degree

180 Degree

 

Frequency

Incidence

Incidence

Incidence

Incidence

Incidence

Incidence

Incidence

Random

in Hz

Angle

Angle

Angle

Angle

Angle

Angle

Angle

Field

1000

0.2

0.1

0.0

0.0

0.0

0.0

-0.2

0.0

1250

0.3

0.1

0.1

0.1

0.1

0.0

-0.5

0.1

1600

-0.1

0.0

-0.1

-0.1

-0.1

-0.1

0.0

-0.1

2000

-0.3

-0.2

-0.2

0.0

0.0

0.0

0.1

-0.1

2500

-0.1

-0.1

-0.2

-0.2

-0.2

-0.1

0.0

-0.2

3150

0.1

-0.2

-0.3

-0.3

-0.3

-0.2

-0.2

-0.3

4000

0.5

-0.2

-0.4

-0.5

-0.5

-0.4

-0.6

-0.5

5000

1.1

0.0

-0.2

-0.3

-0.4

-0.4

-0.7

-0.3

6300

1.1

0.0

-0.1

0.0

-0.1

-0.2

-0.5

-0.1

8000

1.8

0.1

-0.2

0.3

0.6

0.5

0.0

0.1

10000

3.1

0.3

0.0

0.1

0.7

1.2

0.8

0.1

12500

6.1

0.4

-0.9

0.7

1.6

0.8

0.1

0.4

16000

8.2

1.0

0.9

1.1

1.8

2.0

1.0

0.9

Sound level meter type (sections: 5.1.4, 5.1.7, 7.3, 9.2.1 a, and 9.2.5b)

Class 1 when using the BK4936 microphone. Class 2 when using the QE7052 microphone. Group X. Model SE-400 Series. This is applicable with all accessory configurations. The configuration for normal mode of operation with the SE-400 Series is operating the instrument without the remote cable and windscreen. To conform to the specifications for directional response and frequency weightings, the instrument is mounted in a free-field facility using strings.

General Information (section 9.3)

The reference sound pressure level is 114 dB. The one range is the reference range. The microphone reference point is the center of the microphone face.

Sound Examiner SE-400 Series

Page 87
Image 87
3M SE-400 user manual Directional windscreen corrections section, Electrostatic discharges .5.2, 9.2.7 b