62

Data Addendum

Pressure to free field corrections with BK4936 microphone and remote preamp and windscreen (sections: 5.2.8, 9.2.7d)

Note: add to pressure response to get 0incidence free-field response. B&K 4226 calibrator may be used to determine the pressure field response.

 

BK4936 microphone

 

Pressure Field to

 

Free Field

Frequency

Corrections

in Hz

in dB

32

0.1

40

0.0

50

-0.1

63

-0.1

80

-0.1

100

-0.1

125

-0.1

160

-0.1

200

-0.1

250

0.0

315

0.0

400

0.0

500

0.1

630

0.0

800

-0.1

1000

0.0

1250

0.3

1600

0.2

2000

0.5

2500

0.9

3150

1.4

4000

1.7

5000

2.4

6300

3.1

8000

3.8

10000

5.2

12500

6.2

16000

7.8

Sound Examiner SE-400 Series

Page 69
Image 69