79

Data Addendum

Windscreen corrections with QE7052 microphone

*Note: add to the windscreen response to calculate the response without the windscreen

Frequency

Corrections

in Hz

in dB

1000

0.0

1250

-0.2

1600

-0.2

2000

-0.3

2500

-0.4

3150

-0.4

4000

-0.2

5000

0.0

6300

0.0

8000

0.0

10000

0.1

12500

0.2

16000

0.5

Display and Integration (sections: 5.15.5, 5.15.6, 5.15.7, 9.2.5f)

The display update rate is 1 second. The display is updated every second during integration.

Resetting Overload, Under Range, Maximum, and Peak (section 9.2.5 k)

Press the run key to reset. This will reset the measurement data and the indicators. The time to reset is immediate.

Minimum/Maximum integration time (sections: 5.17.1, 5.17.2)

The Minimum/Maximum integration time for measurement time-average levels.

Measurement type

Time average levels

Minimum

1 second

Maximum

99 hours

Reference direction (section: 9.2.5a)

The reference direction is at 0for both BK4936 and QE7052 microphone. The equivalent random incidence direction is 70.

AC/DC Output Characteristics (section 5.16.1)

AC Output:

Weighting: Z-Weighting

Range of Signal: +/- 2.5 Vp

Output Impedance: 2K Ohm

Recommended load impedance: > 40K Ohms

DC Output:

Weighting: A, C, or Z (as set by user)

Range of Signal: 0 to 3.3Vdc, 1mV = 0.1dB (i.e. 450mV = 45.0dB), Under-range ~ 0Vdc, Overload >= 2.0Vdc

Output Impedance: 1K Ohm

Recommended load impedance: > 20K

Sound Examiner SE-400 Series

Page 86
Image 86
3M SE-400 user manual AC/DC Output Characteristics section, Resetting Overload, Under Range, Maximum, and Peak .2.5 k