64

Data Addendum

Pressure to free field corrections with BK4936 microphone and remote preamp and RICR (sections: 5.2.8, 9.2.7d) Note: add to pressure response to get 0incidence free-field response. B&K 4226 calibrator may be used to determine the pressure field response.

 

BK4936 microphone

 

Pressure Field to

 

Free Field

Frequency

Corrections

in Hz

in dB

32

0.1

40

0.0

50

-0.1

63

-0.1

80

-0.1

100

-0.1

125

-0.1

160

-0.1

200

-0.3

250

-0.1

315

0.0

400

0.0

500

0.0

630

0.0

800

0.1

1000

0.0

1250

0.4

1600

0.1

2000

0.4

2500

0.9

3150

1.4

4000

2.0

5000

3.4

6300

4.8

8000

6.4

10000

9.0

12500

10.5

16000

10.8

Sound Examiner SE-400 Series

Page 71
Image 71
3M SE-400 user manual Data Addendum