56

Data Addendum

Pressure to free field corrections with BK4936 microphone (sections: 5.2.8, 9.2.7d)

Note: add to pressure response to get 0incidence free-field response. B&K 4226 calibrator may be used to determine the pressure field response.

 

BK4936 microphone

 

PressureField to

 

Free Field

Frequency

Corctions

in Hz

n dB

32

0.1

40

0.0

50

-0.1

63

-0.1

80

-0.1

100

-0.1

125

-0.1

160

-0.1

200

-0.2

250

-0.1

315

0.0

400

0.1

500

0.2

630

0.1

800

0.0

1000

0.0

1250

0.0

1600

0.2

2000

0.4

2500

0.5

3150

0.8

4000

1.6

5000

2.1

6300

3.1

8000

3.9

10000

5.3

12500

6.2

16000

8.4

Sound Examiner SE-400 Series

Page 63
Image 63
3M SE-400 user manual BK4936 microphone