
5
Turning on
Softkeys
Turning on
To turn on the | key until the start screen appears. Note: the |
Softkeys |
|
The softkey menus provide additional/extension menus on the display and may be available in the measurement, setup, and calibration screens. See Table
Softkeys Explanation
|
|
| Setup icon – The setup icon is used to access the main menu options including: unit information, settings, |
| calibration, memory, |
FS | Fast or Slow Response Time - The response time setting determines how quickly the unit responds to fluctuating |
| noise. Typically, noise is not constant. If you were to try to read the sound level without a response time, the readings |
| would fluctuate so much that determining the actual level would be extremely difficult. While the terms slow and fast |
| have very specific meanings (time constraints), they work very much as you would expect. The fast response would |
| result in a more fluctuating sound level reading than would the slow response. |
ACZ | A, C, or Z Frequency Weighting - These are frequency filters that approximate the equal loudness response of |
| human hearing at low, medium, and high SPL’s. A frequency weighting is the most commonly used filter defined in |
| the international standard IEC 61672:2003. Common applications of A weighting include industrial noise applications |
| and community noise regulations. The A frequency weighted filter makes the sound level meter respond closer to the |
| way the human ear responds to noise at lower levels. It attenuates the low frequency noise below several hundred |
| Hertz as well as the high frequency above six thousand Hertz. C frequency weighting is intended to represent how |
| the ear responds to very high levels. Z weighting is a type of flat response. |
Printer icon - is used for an
CAL | CAL – Cal, or Calibration screen, is used for calibrating the |
LOCK | TLOCK – Shortcut to the secure modes screen. Enables and disables secure and setup modes. |
| Up Arrow indicator: Used in the calibration screen to increase the calibration value. |
| Down arrow indicator: Used in the calibration screen to decrease the calibration value. |
Table
Screen indicators
The screen indicators, located at the top of the display, are a notification tool to alert the user of operational functions. See Table
|
|
|
|
|
Indicators | Explanation | |||
|
|
|
| Secure/Lock indicator - appears if security for run or setup is enabled. |
|
|
|
|
|
|
|
| OL | Overload indicator – appears whenever the level measurement exceeds the range of the instrument. |
|
|
| UR | UnderRange indicator– appears whenever a level measurement is below the range of the instrument. |
|
|
|
| Run indicator- appears when a session is running (measuring integrating measurements.) |
|
|
|
|
|
|
|
|
| Stop indicator- appears when the current session is stopped. |
| 00:00:00 |
| ||
| Run Time: displays the current elapsed logged session run time in hours, minutes, and seconds. | |||
|
|
| / |
|
| Battery charge status- blinks when the battery is low and needs charging. The battery icon shows the | |||
|
|
| ||
|
|
|
| status/level of the internal battery. A full battery indicates the |
|
|
|
| |
|
|
|
| shaded battery icon indicates the battery power is low. When charging, the battery icon will appear as a |
|
|
|
| scrolling, shaded battery charge icon. |
Table
Sound Examiner