Command List

Command

 

 

 

 

Definition

Notes

@

Start / stop continuous data transmission.

 

A

Same as

 

 

 

key.

Key command

RESET

 

 

 

 

 

 

 

 

 

 

D

Set a known tare weight.

“D,1.23CRLF” sets the tare

weight as "1.23kg".

 

 

 

 

 

 

 

 

 

E

Store the unit weight and other values in use to

Refer to the data format.

ID memory.

 

 

 

 

 

 

Refer f-01-05

 

 

 

 

 

 

 

F

Recall a unit weight from ID memory.

“F12CRLF” recalls from ID12.

G

Set a known unit weight.

“G,0.123CRLF” sets the unit

weight as "0.123g" (or "0.123 lb").

 

 

 

 

 

 

 

 

 

J

Same as the

 

 

 

 

 

key.

Key command

TOTAL

 

 

 

 

 

 

 

K

Same as the

 

key.

Key command

M+

 

 

 

 

 

Q

Send data immediately.

Data depends on f-06-02

S

Send stable data after accepting command.

 

T

Same as the

 

 

 

 

key.

Key command

TARE

 

 

 

 

X

Request a list of the F-Function parameters.

The last data terminates with

Y

Request a list of the ID memory contents.

<EOT> (04H)

Z

Same as the

 

 

 

 

key.

Key command

ZERO

 

 

 

ON

Start the scale from power on sequence

 

?ID

Send the ID number in use.

 

?QT

Send the pcs (count) data.

 

?WT

Send the weight data.

Refer to the data format for the

?UW

Send the unit weight in use.

reply.

?AQ

Send the total (accumulated) M+ memory count

 

d t

 

 

 

 

 

 

 

?AN

Send the number of additions to M+ memory.

 

?TR

Send the tare weight in use.

 

?MR

Send the specified ID memory contents.

 

MR

Store the unit weight and tare weight into the

 

specified ID memory.

Refer to the data format for the

 

 

 

reply.

ML

Store the comparator limits into the specified ID

memory.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CM

Clear the specified ID memory contents

“CM,1.2CRLF” clears content of

id12.

 

 

 

 

 

 

 

 

 

?FC

Send the specified F-Function setting.

Refer to the data format for the

FC

Store the specified F-Function setting value.

reply.

Acknowledgment and Error Codes

When the scale receives an external command, it reacts as follows:

‰If the command requests a data reply, the scale will send the data.

For other commands, the scale will send an acknowledgment <ACK><CR><LF> or <ACK> (see F-Function “f-09-02”) upon acceptance of the command.

47

Page 49
Image 49
A&D HC-3Ki, HC-15Ki, HC-30Ki, HC-6Ki manual Command List, Acknowledgment and Error Codes