If you know the exact weight value, or

if you wish to use a different weight,

use the 0 Æ 9 and . 10-key pad to display the desired calibration weight.

(For example: Using 5 kg calibration weight that actually weighs 5.001 kg. Do not forget to enter a decimal point in this case.)

COUNT

888Cal10

WEIGHT

UNIT WEIGHT

Cal 0

5001.

3.Press the ENTER key.

The calibration weight stops blinking.

ENTER

COUNT

888Cal10

 

 

WEIGHT

 

UNIT WEIGHT

 

 

Cal 0

 

 

 

5000.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.Making sure that there is nothing on, or touching the weighing pan, press the ENTER key. When zero calibration is completed, the display will show “Cal f”.

If you don’t need span calibration, press the ON/OFF key to exit from the calibration procedure.

ENTER

Nothing on the

weighing pan!

 

COUNT

888Cal10

 

 

WEIGHT

 

UNIT WEIGHT

 

 

Cal f

 

 

 

5000.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5.Place the calibration weight on the weighing pan and press the ENTER key.

When span calibration is completed, the display returns to step 1 showing the weight value for the calibration weight.

Remove the calibration weight.

If the calibration weight is not what it should be, an error will be displayed. Check if the weight is correct and try again.

6.Press the ON/OFF key to turn the scale off and re-attach the calibration switch cover.

(End of the calibration procedure.)

ENTER Place

Calibration

i ht

COUNT

888Cal02

WEIGHT

UNIT WEIGHT

5000.

If the scale will be moved to another place, set the gravity acceleration value before calibration. The value must be of the area where the calibration is to be done.

28