‰The HC-iis designated as DCE (Data Communication Equipment).

13-3. Data Output Mode

‰The Data Output Modes and Parameters are set by F-Functions in F-06-X as described in the “9-2. F-Functions”

‰To control the scale using commands form an external device, see “13-6. Command Mode”.

‰Refer to “13-6. Command Mode” about the output data format.

Data Output Mode (f-06-01)

‰Key Mode (f-06-01="0")

When the weight display is stable, data is sent by pressing the count display will blink when the data has been sent.

PRINT

key. The

‰Stream Mode (f-06-01="1")

Data is sent continuously. The data-update rate is approximately 10 times per second for f-06-03="2". For f-06-3="0" or "1", the interval between continuous data is approximately 2 seconds.

‰Auto-print Mode A (f-06-01="2")

Data is sent if the weight display is stable at +5d (weighing display division) and above. The next transmission can not occur until after the weight display falls below +5d.

‰Auto-print Mode B (f-06-01="3")

Data is sent if the weight display is stable at ±5d (weighing display division) and above/below. The next transmission can not occur until after the weight display falls between –5d and +5d.

To use with the UFC format, refer to "13-7. Using UFC (Universal Flex Coms) Function"

Data to be Sent (f-06-02)

Select which data to be sent by keying in a (count), weight or unit weight.

0

or

1

for the data: ID No., PCS

Example: Key in 1 1 0 0 to display 1100, this setting would send only the ID number and the count.

8f-06-02

1100

ID No.

PCS (count)

‘0’ not to send data

‘1’ to send data

Data Format (f-06-03)

subtr acts

‰Format for AD-8121 MODE 1 or 2. (f-06-03="0")

‰Format for AD-8121 MODE 3. (f-06-03="1")

‰Format for general apparatuses, computers, etc. (f-06-03="2")